Q: How do you determine if the probe or monitor is faulty?
A: Follow this procedure:
1. Put the #8 switch in the ON position. All other switches must
be OFF.
2. Dry out the probe and suspend it in mid-air.
3. Depress the push-to-test button.
3A. If the GREEN light is ON, immerse the probe in tap water.
Depress the push-to-test button.
If the GREEN light is ON, you have a bad monitor and/or
bad
probe.
If the RED light is ON, you have a bad cartridge and/or
bad
membrane.
OR
3B. If the RED light is ON, unplug the probe from the jack (Fig. D).
Depress the spring contact on that jack with a pencil tip so
that it is no longer touching the main body of the jack.
Depress the push-to-test button.
If the GREEN light is ON, you have a bad probe.
If the RED light is ON, you have a bad monitor and/or
bad
probe.
4. Be sure to restore the monitor to the #7 switch ON and all others
OFF.
IMPORTANT NOTE: ONLY USE ALKALINE BATTERIES.
Other Batteries will cause erroneous readings
Monitor Testing
Fig. D: Purity Monitor Jack
Depress the
spring contact
SpectraPure
®
SpectraPure®Inc
.
Fax 480.894.6109 Fax us toll-free 1.877.527.7873
E-mail: [email protected] Visit us on the web www.spectrapure.com
3
System Information
With this configuration, the possibility of contaminated product
water due to exhausted cartridges is completely eliminated. With
the majority of contaminants now removed by Stages #1 and
#2, the water is then passed through two additional stages of
DI resin. Stages #3 and #4 have our Chloramine Removal Car-
tridges (DI-AR-CI-10). They will provide longer life than standard
mixed-bed cartridges, when used to treat water containing
chlorine, chloramine and ammonia . These two additional stages
of purification polish the water, adjust the pH to neutral, and
remove any remaining contaminants to achieve greater than 18
megohm purity.
An electronic purity monitor has been included in order to pro-
vide accurate monitoring of the condition of stage #3. Stage
#4 acts as a backup cartridge, which ensures that, if Stage #3
becomes exhausted, no contaminaton of the product water will
result.
Stages #3 and #4 utilize the same scheme of rotating cartridges.
These cartridges will also change color. With this superior con-
figuration, you can count on the best quality water possible,
combined with the longest cartridge life available.
The Ultimate™ is also available as a system optimized for labora-
tory water sources. This model has all the advantages of the
standard Ultimate™ System, except Stages #3 and #4 are con-
figured with our Semi-Conductor Grade Mixed-Bed Cartridges
(DI-MB-10).
Any existing RO or RO/DI System can be converted to The
Ultimate™. Contact the factory for further details.