RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND
MOBILE RF COMMUNICATIONS EQUIPMENT
AND THE PORTABLE ELECTRICAL STIMULATOR UNIT
This Aspire 2 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of this Aspire 2 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and this Aspire 2 as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output
Separation distance according to frequency of transmitter
power of transmitter
m
W
150 KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance ’d’
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where ’P’ is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is afected by
absorption and refection from structures, objects and people.
d
=
1.2
d
=
d
=
Returning
Equipment
for
Repair
Be sure to call for an authorization number (RA) before returning any equipment.
Send the unit(s) postage prepaid and insured, with proof of purchase, to the
address below. If you are shipping from outside the United States, mark the
package ‘Goods to be repaired-Made in China’ to avoid unnecessary customs
charges.
Provide a detailed description of the problem you are experiencing, along with your email,
phone number and fax number if applicable.
Technical
Support
and
Repair
1.2
2.3
Any queries should be addressed to Spectramed at:275 W Johnstown Rd
Gahanna, OH 43230