Setting Threshold

Patients will demonstrate varied levels of EMG activity and thus, the effects of 
doing ETS may be different between patients. Optimal effects can be achieved 
if the user downloads an appropriate threshold from the APP.

Download “0” from the iPad, the digital display will read AUTO in the center of 
the yellow area.

1) Press

2) Stay relaxed for 6 to 8 seconds.

3) Stay flexed for 8 to 10 seconds. The stimulator will then set the threshold.

4.3 Other Functions

When battery power is lower than 20%, the battery icon on the OLED will flash 
for 1 second to promote recharging; when battery is lower than 10%, the icon 
will flash for 0.5 second and electrical stimulation will shut off; when battery is 
lower than 5%, the stimulator will automatically shut down.

4.3.1 Low battery warning

4.2.3 GAME

GAME  is  the  training  mode  that  combines  sEMG  biofeedback  treatment  with 
interactive  gaming  technology.  Biofeedback  is  the  use  of  sEMG  signals  to 
provide  a  real  time  visual  display  in  ether  a  graph  or  animated  format  of  the 
muscular effort exerted by the patient. This external feedback helps the patient 
gain greater awareness of the effort exerted and gives them a tangible goal to 
work towards thereby improving their performance and affecting a physiological 
change.  The  animations  provide  an  enjoyable,  engaging  format  during  the 
rehabilitation  process.  Different  games  have  been  designed  specifically  to 
address 3 different types of muscle function; strength, endurance and coordina-
tion. The system transfers the collected EMG signal to the iPad through Blue-
tooth technology to control the games.  

mA

23

.0

mA

23

.0

mA

0

.0

ETS

ETS

500

AUTO

500

AUTO

Relax:

1

500

AUTO

Relax:

OK!

500

AUTO

Flex:

3

500

AUTO

Flex:

OK!

Then press 

   

, and the stimulator starts to work. Please see following interface:

ETS

 

Mode

Start to work

The initial current is 0 mA for security. User can press        or         to adjust the 
intensity.

ETS

500

ETS

500

mA

0

.0

1. Press      or      to switch to the fifth prescription, press       to enter the setting page. 
2. Press     or      to select parameters, below are the adjustable parameters: 
RQ: frequency 
PD: pulse width 
CT: duration of stimulation 
RT: rest time between 2 stimulations 
RU: ramp up 
RD: ramp down 
Times: Cycles of stimulation
3. Press      or       to adjust the parameters.
4. Press      to confirm.  

 

Manually set the customized prescription

Note: If user downloads the program from iPad, the fifth program will display as
P1 to P42 If user customizes the program by the stimulator, the fifth program will 
display as C.

*

Содержание 01AS80

Страница 1: ...m Date 2016 12 23 sEMG Biofeedback FES Neuro Muscular Electrical Stimulator 01AS80 User manual Thanks for buying our product Please carefully read the user manual before use and keep the manual in a safe place ...

Страница 2: ...ions Skin Care Chapter 3 The Aspire 2 Description Features Settings Chapter 4 Operation Power on Mode Selection Other Functions Chapter 5 Care and Maintenance Stimulator Battery Lead Wires Guardian Self Adhesive Electrodes Chapter 6 Specifications and Parameters Chapter 7 Manufacturer s Declaration Warranty Customer Service 01 01 01 02 04 04 04 04 05 06 06 07 07 07 08 09 09 09 13 15 15 15 16 16 17...

Страница 3: ...rehabilitation treatment programs for each patient 1 2 Main Innovations 1 Technology of self EMG signal acquisition and processing 2 New generation of real time biofeedback functional electrical stimulation technology based on EMG signal and muscle strength 3 iPad interactive games training and rehabilitation therapy applications APP 4 Multiple rehabilitation training modes in the APP Warranty Spe...

Страница 4: ...It is customer responsibility to assure that this equipment and vicinity equipment complies the value of radio frequency inter ferences shown in General Regulation for safety according to IEC 60601 1 2 Tables as described in this section The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifcations to ...

Страница 5: ... 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequenc...

Страница 6: ...ld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY This Aspire 2 is intended for use in the electromagnetic environment specified below The customer or the u...

Страница 7: ... 2 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The Aspire 2 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings and homes used for domestic purposes MANUFACTU...

Страница 8: ...atio 100dB NMES Frequency 5 100Hz biphasic balanced wave 10 Pulse width 50 450uS 10 Output intensity load 1000Ω min 1mA max 60mA ETS Frequency 18 Hz biphasic balanced wave 10 Pulse width 150 uS 10 Output intensity load 1000Ω min 1mA max 60mA Working Environment Temperature 5 40 Relative humidity 80 HR Atmos 86Kpa 106Kpa Storage Environment Temperature 20 55 Relative humidity 93 HR Atmos 70Kpa 106K...

Страница 9: ...wires carefully after each use Guardian Self Adhesive Electrodes Check the short connectors to make sure they have not become separated from the electrodes Replace electrodes onto plastic film after use If they drop onto the floor debris will adhere to conductive gel rendering the electrodes ineffective Optimal electrode function cannot be guaranteed after 4 uses Keep electrodes away from water Tr...

Страница 10: ...o not use cleaning sprays or alcohol based cleaning solutions Do not store in direct sunlight in high temperatures moist areas dusty areas or near corrosive gas Do not immerse in water Do not throw step on or exert pressure on the stimulator When the stimulator reaches the end of its service life 3 years please dispose of it in accordance with the local and national regulations Battery The Aspire2...

Страница 11: ... up the stimulator will shut down and cannot work again until downloading another treatment time from the APP 4 3 5 Current Limit A current limit value can be downloaded from the APP to limit the maximum output current The available working current of the stimulator is 0 to 60 mA however if the user downloads a current limit value of 25 mA from the APP then the maximum output current will stop at ...

Страница 12: ...e effort exerted and gives them a tangible goal to work towards thereby improving their performance and affecting a physiological change The animations provide an enjoyable engaging format during the rehabilitation process Different games have been designed specifically to address 3 different types of muscle function strength endurance and coordina tion The system transfers the collected EMG signa...

Страница 13: ...r muscle movements as the reliability of the signal will be directly impacted by the type of swallow elicited such as dry swallow 5cc water 10cc water puree or solid bolus Proper use of sEMG can be very useful in providing objective data for establishing baseline and progress over time Improper use of sEMG will not yield useful or accurate data The stimulator has 5 prescriptions as follows 4 fixed...

Отзывы: