4.2.2 NMES

Neuromuscular  Electrical  Stimulation  (NMES)  is  an  electrical  stimulation  training 
mode.  NMES  provides  an  electrical  impulse  through  surface  electrodes  to  the 
peripheral  nerves  to  evoke  an  action  potential.  Repeated  assisted  contractions 
of a weak, dysfunctional neuromuscular system results in stronger, more efficient 
and coordinated muscle movement. The iPad APP has 4 fixed prescriptions and 
1  customized  prescription  (Custom  settings  by  medical  professionals  according 
to  patients'  actual  conditions)  for  different  body  parts  based  on  international 
common standards of electrical stimulation and our company's independent study.

sEMG TEST

The sEMG test in the APP detects the patient's EMG signal and provides a sum 
of the activity measured in microvolts and a display of the results. This objective 
data can be used to establish baseline and progress over time. The threshold is 
set in “Setting” by comparing the test results.
After setting, the threshold will be recorded by the system. If user doesn't set 
threshold, the system will work with the previously set threshold. 

Swallowing sEMG TEST

The swallowing sEMG test detects the EMG signal during a swallow and provi-
des  a  visual  display  of  the  average  microvolts.  sEMG  for  swallowing  is  very 
different than sEMG for other muscle movements as the reliability of the signal 
will be directly impacted by the type of swallow elicited such as; dry swallow, 
5cc water, 10cc water, puree or solid bolus. Proper use of sEMG can be very 
useful in providing objective data for establishing baseline and progress over 
time. Improper use of sEMG will not yield useful or accurate data.

The stimulator has 5 prescriptions as follows:
4 fixed prescriptions and 1 customized which can either be downloaded from 
the APP or manually set through the stimulator.

The following are the 4 fixed prescriptions, each one lasts about 30 minutes:

After selecting a prescription, press         to start training.

The initial current is 0 mA for security. User can press         or        to adjust the 
intensity.

Proper use of NMES requires that the stimulus is modified and manipulated 
according to the patient's individual needs, the 4 prescriptions provided may 
work  for  the  majority  of  patients  but  in  many  cases  the  stimulus  must  be 
customized for an optimal response. A custom prescription can be created 
by  connecting  the  stimulator  to  iPad,  clicking  on  the  5th  prescription  and 
entering the parameters needed.

 Load customized prescription

-A1

NMES

-A2

NMES

-A3

NMES

-A4

NMES

-P4

NMES

-A1

14.5mA

NT:  52

0

RT: 

8

00

-A1

14.5mA

NT:  52

0

RT: 

8

00

-A1

NMES

Fixed Prescription

Customized Prescription

4) After the threshold is set, the digital display will show the current and the 
    user can increase to the appropriate current level and start the training 
    process.

mA

23

.0

mA

23

.1

*

Содержание 01AS80

Страница 1: ...m Date 2016 12 23 sEMG Biofeedback FES Neuro Muscular Electrical Stimulator 01AS80 User manual Thanks for buying our product Please carefully read the user manual before use and keep the manual in a safe place ...

Страница 2: ...ions Skin Care Chapter 3 The Aspire 2 Description Features Settings Chapter 4 Operation Power on Mode Selection Other Functions Chapter 5 Care and Maintenance Stimulator Battery Lead Wires Guardian Self Adhesive Electrodes Chapter 6 Specifications and Parameters Chapter 7 Manufacturer s Declaration Warranty Customer Service 01 01 01 02 04 04 04 04 05 06 06 07 07 07 08 09 09 09 13 15 15 15 16 16 17...

Страница 3: ...rehabilitation treatment programs for each patient 1 2 Main Innovations 1 Technology of self EMG signal acquisition and processing 2 New generation of real time biofeedback functional electrical stimulation technology based on EMG signal and muscle strength 3 iPad interactive games training and rehabilitation therapy applications APP 4 Multiple rehabilitation training modes in the APP Warranty Spe...

Страница 4: ...It is customer responsibility to assure that this equipment and vicinity equipment complies the value of radio frequency inter ferences shown in General Regulation for safety according to IEC 60601 1 2 Tables as described in this section The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifcations to ...

Страница 5: ... 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequenc...

Страница 6: ...ld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY This Aspire 2 is intended for use in the electromagnetic environment specified below The customer or the u...

Страница 7: ... 2 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The Aspire 2 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings and homes used for domestic purposes MANUFACTU...

Страница 8: ...atio 100dB NMES Frequency 5 100Hz biphasic balanced wave 10 Pulse width 50 450uS 10 Output intensity load 1000Ω min 1mA max 60mA ETS Frequency 18 Hz biphasic balanced wave 10 Pulse width 150 uS 10 Output intensity load 1000Ω min 1mA max 60mA Working Environment Temperature 5 40 Relative humidity 80 HR Atmos 86Kpa 106Kpa Storage Environment Temperature 20 55 Relative humidity 93 HR Atmos 70Kpa 106K...

Страница 9: ...wires carefully after each use Guardian Self Adhesive Electrodes Check the short connectors to make sure they have not become separated from the electrodes Replace electrodes onto plastic film after use If they drop onto the floor debris will adhere to conductive gel rendering the electrodes ineffective Optimal electrode function cannot be guaranteed after 4 uses Keep electrodes away from water Tr...

Страница 10: ...o not use cleaning sprays or alcohol based cleaning solutions Do not store in direct sunlight in high temperatures moist areas dusty areas or near corrosive gas Do not immerse in water Do not throw step on or exert pressure on the stimulator When the stimulator reaches the end of its service life 3 years please dispose of it in accordance with the local and national regulations Battery The Aspire2...

Страница 11: ... up the stimulator will shut down and cannot work again until downloading another treatment time from the APP 4 3 5 Current Limit A current limit value can be downloaded from the APP to limit the maximum output current The available working current of the stimulator is 0 to 60 mA however if the user downloads a current limit value of 25 mA from the APP then the maximum output current will stop at ...

Страница 12: ...e effort exerted and gives them a tangible goal to work towards thereby improving their performance and affecting a physiological change The animations provide an enjoyable engaging format during the rehabilitation process Different games have been designed specifically to address 3 different types of muscle function strength endurance and coordina tion The system transfers the collected EMG signa...

Страница 13: ...r muscle movements as the reliability of the signal will be directly impacted by the type of swallow elicited such as dry swallow 5cc water 10cc water puree or solid bolus Proper use of sEMG can be very useful in providing objective data for establishing baseline and progress over time Improper use of sEMG will not yield useful or accurate data The stimulator has 5 prescriptions as follows 4 fixed...

Отзывы: