![Spectracom Pendulum PathAlign-R 2200 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/spectracom/pendulum-pathalign-r-2200/pendulum-pathalign-r-2200_user-manual_1343634031.webp)
Spectracom Corporation
PathAlign-R
2200 / 2240 / 2241 PathAlign-R Manual
3-23
3.3
Rear Panel
Figure 3-2 depicts the PathAlign-R rear panel connectors and battery across door.
Figure 3-2. Model 2240 Rear Panel
Name Function
Battery Access Door Hinged battery access door provides for removal/replacement
of internal battery. Door latches via a thumbscrew.
Osc. Calib. Access hole allowing the use of a trim pot adjusting tool to fine-
tune the internal oscillator during yearly calibration.
Note:
Refer oscillator calibration to a qualified technician!
External Charger IN Type 5.5 mm x 2.1 mm, two-terminal battery power connector
connects the external 15VDC/ 1A Battery Charger to the
PathAlign-R (center pin is positive, maximum charge current: 1
amp).
Remote Readout BNC type connector. Provides a 0-2 VDC output that tracks the
front panel LCD Path Loss Meter (1 mV = 0.1 dB). Allows for
external readout of path loss using a handheld DVM.
3.4
Operating Characteristics
The following paragraphs describe the operating range, band resolution, dynamic range, and
overload characteristics.
3.4.1 Operating Ranges
There are seven operating ranges available: Band 1: 1.5 GHz to 2.5 GHz; Band 2: 3.5 GHz to
5.0 GHz and 5.8 GHz to 6.6 GHz; Band 3: 7.5 GHz to 10.0 GHz and 11.0 to 12.0 GHz; Band 4: