Installation
2.1
Introduction
This chapter outlines the procedure for initial inspection and installation of the PathAlign-R. The
subject matter includes receiving inspection, power requirements, battery information,
installation criteria, mating connectors, and incoming confidence check. Instructions for
preparing an instrument for reshipment and storage instructions are also included.
2.2
Receiving Inspection
Prior to accepting the unit from the shipper, inspect the condition of the shipping container for
any evidence of freight damage. Any damage should be noted by both the shipper and the
receiving customer and reported to the insurance investigator.
Immediately after removing the instrument from the shipping carton, inspect it for any possible
physical damage that may have been incurred in shipment. Check for surface scratches and
dents; note the condition of switches, buttons, and connectors (carrying bag, battery, cables,
etc.)
Should any damaged be detected, notify Spectracom. DO NOT USE THE
INSTRUMENT UNTIL INSTRUCTED TO DO SO BY THE FACTORY.
2.3
Power Requirements
The PathAlign-R is powered by a 12VDC/2.3 Ah rechargeable battery which can be removed
via the rear panel Battery Door. Continuous Battery operating time is approximately 4-5 hours
@ 25°C (77°F). A ’Low Battery’ indicator is contained within the LCD display and is displayed
when the battery has approximately 15 minutes of operations remaining. Charging of the battery
while inside of the PathAlign-R is accomplished by plugging the battery charger’s DC plug into
the PathAlign-R’s rear panel ’External Charger In’ jack (charging time is approximately 3-hours
to fully charge the battery). The PathAlign-R Battery Charger operates, when attached to an
appropriate AC MAINS source, over an AC line voltage of 90 VAC to 264 VAC, 47 Hz to 63 Hz.
The charger’s detachable AC MAINS three-conductor power cord (type IEC 320-3:1987/EN 60
320) conforms to the CEE Color Coding Standards (Light Blue: Neutral; Brown: Live;
Green/Yellow: Protective Ground) and grounds the instrument’s chassis when connected
through a compatible grounded 3-wire power receptacle.
Note: Use only the Battery Charger supplied with the
PathAlign-R, another charger may damage the
PathAlign-R or the battery!
If the charger is lost or
damaged, a replacement may be purchased from