POSSIBLE PROBLEM
SOLUTION
Low Suction (cont.)
Silicone valve is not attached
correctly.
Reposition the valve and then
reassemble.
Backflow protectors are not
assembled correctly.
Reassemble and reattach the
backflow protectors.
Tubing is not fully attached.
Firmly attach tubing to both the
backflow protector and the pump
nozzle.
If single pumping, nozzle cover
is loose.
Ensure the second pump nozzle is
fully covered by the rubber cap.
Air is leaking out between
breast and the breast shield.
Adjust breast. If persists, breast
shield may be the wrong size.
Valves and backflow protector
may need replacing.
Over time the elasticity of these
parts degrade which impacts
suction. Replace regularly.