Spectra 10K Library Preventative Maintenance Guide

8.

Select 

Display DSP Table

 to review the test results.

Pay close attention to the Test Status Column—the first column on the left.

The first four rows should be ignored.

All other values in this column should be either 

0

 or 

2

.

Note: 

If any number in the first column (after the first four rows) is either 

1

 or a 

3

, contact Spectra Logic Technical Support at (800) 227-4637 

or (303) 449-0160, or send an e-mail to [email protected].

9.

Touch anywhere in the screen to exit the DSP trace window.

10.

Select the 

Exit

 icon and then select 

Reset

 to return the unit to a known condition.

11.

Remove the scratch tapes from the library.

12.

Use your backup software to re-insert your data tapes into the library.

13.

Verify that all drives respond to server commands and that the library boots up 
correctly.

Test Status Column

Error Code Buffer Column

Содержание 10K Library

Страница 1: ...Spectra 10K Library Preventative Maintenance Guide SDOC 2001 1 Revision C Copyright Spectra Logic Corporation All rights reserved...

Страница 2: ...eventative Maintenance Guide Contents About Preventative Maintenance 3 Preparing the Library 4 Changing the Air Filter 5 Cleaning the Door Sensor 6 Internal Maintenance and Cleaning 7 Cleaning the Dri...

Страница 3: ...e Spectra 10K library Time Involved This procedure takes approximately two hours Tools Needed A credit card or thin piece of plastic A 1 Phillips screwdriver with a long handle 8 inch minimum A large...

Страница 4: ...s were in which slots 2 Shut off all power to the library 3 Shut down the host system and switch off the power 4 Remove the library s AC power cord from the AC power outlet 5 Carefully remove SCSI cab...

Страница 5: ...to open the grill 2 Swing open the grill 3 Remove the air filter 4 Inspect the grill if the grill is dirty clean it 5 Install the new air filter 6 Close the grill and snap it shut Note Filters should...

Страница 6: ...he back of the door and two holes near the front of the door If these holes are clogged with dust or other debris the door sensors will fail to operate 2 Use the can of compressed air to blow out the...

Страница 7: ...n the back of the library Caution The library contains static sensitive devices Please use a static dissipative wrist strap or other ESD equipment and tools for this process 2 Remove the cover by a Sl...

Страница 8: ...pectra 10K Library Preventative Maintenance Guide 4 Slide the drives as far back as they will go without removing the drive cables All drives in original positions All drives slid back Drive 2 slid ba...

Страница 9: ...nnot be completely removed Loosen the screws sufficiently to disengage them from the chassis 6 Rotate the robotic arm so that the gripper jaws point to the rear of the library as shown below the carou...

Страница 10: ...e of the robotic arm c Move the carousel toward the rear of the library so that the front of the carousel touches the robotic arm base d Lift the rear of the carousel above the robotic arm base The ca...

Страница 11: ...y Caution Make sure that the door cable is not pinched between the carousel and the base plate or UCM board see the pictures below 13 Slide the drives back to their original positions next to the caro...

Страница 12: ...e sure that boards are firmly seated on the base plate 3 Check all cable connections including the drive cables and make sure that all cables are securely attached to boards 4 On differential units ch...

Страница 13: ...the cleaning frequency until you reach optimal operation Allow the environment to determine how often you clean the drives each environment is unique This procedure must be performed with the library...

Страница 14: ...n door Insert the Sony AIT cleaning cartridge into the cartridge door and select OK The robot takes the cleaning cartridge from the door and inserts it into the drive When the drive has finished the c...

Страница 15: ...Attach SCSI cable connections to the back of the library 2 Reconnect AC power to the library 3 Turn on power to the library 4 Wait for the library to come to a Unit Ready state Read Write Confidence...

Страница 16: ...16 Spectra 10K Library Preventative Maintenance Guide 3 Select 1 2 3 then select OK 4 Select R W Confidence 5 When prompted place the scratch tape in the door and select OK...

Страница 17: ...lows For AIT 1 the values should be less than 02 00 percent For AIT 2 and AIT 3 the values should be less than 04 00 percent Note If the results on any drive are higher than the percentage listed abov...

Страница 18: ...18 Spectra 10K Library Preventative Maintenance Guide 3 Select Diagnostics 4 Select 1 2 3 then select OK 5 Select Qualification Tests...

Страница 19: ...ectra 10K Library Preventative Maintenance Guide 6 Select Confidence Test After the test completes a window appears that indicates Test Passed or Test Failed 7 Select OK the Qualification screen reapp...

Страница 20: ...is either a 1 or a 3 contact Spectra Logic Technical Support at 800 227 4637 or 303 449 0160 or send an e mail to support spectralogic com 9 Touch anywhere in the screen to exit the DSP trace window...

Отзывы: