![Specac Silver Gate Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/specac/silver-gate/silver-gate_user-manual_1341492015.webp)
Silver Gate™ Evolution ATR System
13
3. Maximize the energy by adjusting the rotation and tilt movement
screws (
8
) and (
9
) respectively of mirror
M1
. (See Figure 6.)
4. Repeat the rotation and tilt adjustments as in steps 2 – 4 until
maximum transmission is achieved and then replace the front
cover plate (
6
) and tighten the cover plate thumb screws (
7
).
5.
If the Benchmark™ baseplate is mounted on a rail or is movable in
the spectrometer, move the Silver Gate™ ATR Accessory on the
Benchmark™ baseplate to find the best position where the
transmission is at a maximum. Now firmly secure the Benchmark™
baseplate in this position.
Note:
In some spectrometers it may be necessary to remove the
Silver Gate™ optical unit before securing the baseplate. Where
this is the case, ensure that the Benchmark™ baseplate does
not move when removing the Silver Gate™ optical unit.