![Specac Silver Gate Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/specac/silver-gate/silver-gate_user-manual_1341492013.webp)
Silver Gate™ Evolution ATR System
11
The flat and recessed ATR crystal top plate assemblies will fit into the
top surface area of the Silver Gate™ accessory in one orientation only.
Align the location holes (
2
) with the location pins (
3
). The location
slot
(
2
) must be towards the front of the Silver Gate™ Accessory where the
thumbscrew fixing (
1
) is located. Allow the three magnetic catches (
4
)
to engage with their three counter plate fixings (
5
) to hold the ATR
crystal top plate securely in position. To remove and swap over with
any other ATR crystal top plate assembly, just pull up on the ATR
crystal top plate to release the magnetic catches.
Alignment
With an ATR crystal top plate assembly in position on the optical unit
and when the whole is installed in the spectrometer on an appropriate
Benchmark™ baseplate, the Silver Gate™ Accessory can be finely
aligned. The optical unit cover plate (
6
) is removed by undoing the
cover plate thumb screws (
7
) (see Figure 5.) to gain access to the
mirror adjustment screws (
8
) and (
9
).
Figure 6. Front view of Silver Gate™ with a Flat ATR Crystal
Top Plate and Optical Unit Cover removed to show Mirrors
8
9
Mirror
M1
Mirror
M2
Содержание Silver Gate
Страница 1: ...Silver Gate Evolution Single Reflection ATR System User Manual 2I 10680 Issue 3 ...
Страница 2: ......
Страница 3: ...Silver Gate Evolution Single Reflection ATR System User Manual 2I 10680 Issue 3 ...
Страница 40: ...User Manual 38 Notes ...
Страница 41: ...Silver Gate Evolution ATR System 39 Notes ...
Страница 42: ...User Manual 40 Notes ...