background image

User Manual 

 

12

Cleaning of Pellet Die Parts and KBr Pellet Removal 
Intact from the Pellet Ring Holder 

 
When the KBr pellet sample being held in the die ring holder (

6

) has 

been analysed, the KBr sample must be removed from the die ring 
holder (

6

) for pressing of a new sample into the same die ring holder. 

 
To remove the KBr pellet, the entire KBr pellet and die ring holder (

6

assembly can be rinsed with cold or warm water to gradually dissolve 
the KBr sample. When the KBr sample has been washed away from 
inside the holder, the die ring holder (

6

) part can be further 

washed/rinsed with methanol and then dried with tissues. Specac 
would also then recommend that this part could be stored in dry 
environment or placed on a heated surface (circa 30°C to 35°C) prior 
to use for the next time.  
 
After initial pressing of the pellet into the die ring holder (

6

) (from 

sample preparation stage), the top anvil (

8

) and bottom anvil (

9

) should 

also be rinsed with water, then methanol, dried with tissues and stored 
on a heated top plate surface ready for next use.  
 

However

, if you would wish to 

save

 the 7mm KBr pellet that has been 

formed within the die ring holder (

6

) after it spectroscopic analysis, 

then this KBr pellet can be removed intact from the die ring holder (

6

part, by use of the top anvil (

8

) with the longer plunger and the Perspex 

die extractor cap piece (

10

). 

 
Take the die ring holder (

6

) with KBr pellet and place the plunger of the 

top anvil (

8

) into the aperture hole on one side of the die ring holder 

(

6

). Place this combined assembly with the Perspex extractor cap (

10

in contact with other side of the die ring holder (

6

) into the Mini-Pellet 

Press pressing area (

2

) as shown in Fig 8. Make sure that this 

assembly of parts is placed centrally between the lead screw pressing 
face (

12

) and the piston surface (

11

). 

Содержание GS03940

Страница 1: ...Mini Pellet Press User Manual 2I 03940 6...

Страница 2: ......

Страница 3: ...Mini Pellet Press User Manual 2I 03940 6...

Страница 4: ...T DIE ASSEMBLY FROM THE MINI PELLET PRESS 10 POSITIONING THE KBR PELLET DIE RING HOLDER IN THE SPECTROMETER 11 CLEANING OF PELLET DIE PARTS AND KBR PELLET REMOVAL INTACT FROM THE PELLET RING HOLDER 12...

Страница 5: ...and compress a sample held in the pressing area To assist in the turning of the pressure screw knob an M8 6mm A F size short handled Allen key is included to fit the central socket screw for greater l...

Страница 6: ...ssembly P N GS03950 has been ordered too check that the following items have been included 7mm Pellet Die assembly P N GS03950 including its own Pellet Ring Holder Top Anvil Bottom Anvil and Perspex E...

Страница 7: ...mm pellet die assembly 1 in position in readiness to turn the pressure screw hand knob assembly 4 Only a light force is needed to turn the pressure hand knob 4 to provide a sufficient tonnage load but...

Страница 8: ...ing mill such as the Specamill P N GS06000 with an agate capsule set P N GS06200 The 7mm pellet die assembly complete 1 consists of a central pellet ring holder 6 top anvil 8 a bottom anvil 9 and a cl...

Страница 9: ...d to compact the powder mixture into the hole The mixture is sufficiently soft to allow the plunger part of the top anvil 8 to fit and locate loosely into place in the aperture hole of the 7mm pellet...

Страница 10: ...ce of the leadscrew 12 and the piston 11 of the Mini Pellet Press The piston 11 will be at its rest position for the start of its travel This will be when there is no applied pressure on the system fr...

Страница 11: ...ly 4 is fully retracted the pressing piston 11 will be at its correct start position for pressing When the Mini Pellet Press pressing area has been set to accept the 7mm pellet die assembly 1 place th...

Страница 12: ...ge reading stop turning the pressure screw knob assembly 4 This will be a sufficient tonnage load to have compacted the KBr powder together to form a suitable pellet Note On the pressure gauge 13 ther...

Страница 13: ...nd The KBr sample powder material will have been compacted into a 7mm diameter pellet contained in the die ring holder 6 central aperture hole The complete die ring holder 6 with KBr pellet is now pla...

Страница 14: ...l pressing of the pellet into the die ring holder 6 from sample preparation stage the top anvil 8 and bottom anvil 9 should also be rinsed with water then methanol dried with tissues and stored on a h...

Страница 15: ...e Perspex extractor cap 10 When the KBr pellet has been pushed out of the die ring holder 6 retract the lead screw assembly 3 by a couple of anticlockwise turns and carefully remove the assembly of pa...

Страница 16: ...omplete P N GS03951 7mm Pellet die ring holder body P N GS03952 7mm Pellet die top anvil long plunger P N GS03953 7mm Pellet die bottom anvil short plunger P N GS03954 7mm Pellet die Perspex extractor...

Страница 17: ...ce 12 Piston 13 Pressure gauge 14 O rings for ring holder fixing when fitted on 3 x 2 slide mount Specifications of the Mini Pellet Press Maximum load 2 tons Pellet Die diameter 7mm Maximum space betw...

Страница 18: ...User Manual 16 Blank Page Notes for Use of Mini Pellet Press...

Страница 19: ...te www systems eng co jp Spain Teknokroma S Coop C Ltda Barcelona Tel 93 674 8800 Fax 93 675 2405 email comercial teknokroma es Switzerland Portmann InstrumentsAG Biel Benken Tel 061 726 6555 Fax 061...

Страница 20: ...PECAC INC 414 Commerce Drive Suite 175 Fort Washington PA 19034 USA Tel 215 793 4044 Fax 215 793 4011 SPECAC LTD Unit 12 Science Innovation Centre Halo Business Park Orpington Kent BR5 3FQ Tel 44 0 16...

Отзывы: