Mini-Pellet Press
11
Positioning the KBr Pellet Die Ring Holder in the
Spectrometer
Having removed the compressed 7mm pellet die assembly (
1
) from the
Mini-Pellet Press, the top anvil (
8
) and bottom anvil (
9
) are
separated/removed from the die ring holder (
6
). Hold the knurled outer
edge of the die ring holder (
6
) with one hand and twist/turn the anvils
(
8
and
9
) by holding their knurled edges with the other hand. The
KBr/sample powder material will have been compacted into a 7mm
diameter pellet contained in the die ring holder (
6
) central aperture
hole.
The complete die ring holder (
6
) with KBr pellet is now placed into an
infra red spectrometer system by use of the 3” x 2” slide mount holder
(P/N GS03960) (
6
). The die ring holder (
6
) rests simply on the two
support rods of the 3” x 2” slide mount holder (
7
) and is held tight
against the 3” x 2” slide plate by compression of the two O-rings (
14
)
supplied having been slid over the support rods as shown. (See Fig 7.)
Fig 7. 7mm Pellet Die Ring Holder on 3” x 2” Slide Mount
7
6
14