Ultraview Digital Telemetry
3-2
071-0774-00 Rev. A
Setting Up SpO
2
Monitoring
The 91343 digital telemetry multi-parameter transmitter uses Spacelabs Medical
sensors as well as those from other manufacturers. Refer to
Accessories
on page
1-12
for information concerning specific sensors.
All sensors require an adapter cable between the sensor and the transmitter.
Because the adaptor cable is reusable, do not discard it when you have finished
using a disposable oximetry sensor. Disconnect the sensor cable from the adapter
cable before discarding the sensor.
To connect the SpO
2
adapter cable to the transmitter, align the cable with the
notch on the front of the transmitter connector, and push the cable straight down
into the transmitter. To remove the cable, press the latch release on the bottom of
the cable, and pull the cable straight out.
Figure 3-1: SpO
2
Adapter Cable to Transmitter
To enable SpO
2
monitoring in the 91343 digital telemetry multi-parameter
transmitter, choose an averaging interval of 4, 8, or 16, seconds by setting DIP
switches 1 and 2 as explained in
Setting SpO2 Data Averaging Period and
Sampling Interval
on page 3-6.
CAUTION:
•
Use only patient sensors specified by Spacelabs Medical.
Using sensors other than those specified may degrade
performance and damage the transmitter during
defibrillation.
•
Check the sensor site frequently. Do not allow the sensor
to remain on one site for a prolonged time, especially when
monitoring neonates. Refer to the sensor manufacturer's
instructions.
•
Never attach an SpO
2
sensor on a limb being monitored
with a blood pressure cuff or a limb with restricted blood
flow.
•
A poorly applied sensor may give inaccurate saturation
values.
•
Choose a site with sufficient perfusion to ensure accurate
oximetry values.
CAUTION:
•
Never twist the cable.
To set up SpO
2
monitoring:
1
Open the battery cover and
remove the battery
2
Confirm that the DIP switches 1
through 8 are in the correct
setting (Switch 7 must be set to
ON for neonatal use and to OFF
for adult use)
3
Reinstall the battery and close
the battery cover
4
Connect the SpO
2
adapter
cable (P/N 700-0014-00) to the
transmitter
5
Attach the sensor to the patient
and connect the sensor cable to
the SpO
2
adapter cable
6
Initiate ECG monitoring
7
Touch ECG
8
Touch CHANNEL FORMAT
9
Select SpO2 ON