Ultraview Digital Telemetry
2-6
071-0774-00 Rev. A
Restoring Default Settings
With the Module Configuration Manager feature, you can restore all default
settings. User-configurable options are listed in the
Module Configuration
Manager
chapter of the
UCN Operations Manual
.
Changing the Display Resolution
The MONITOR/EXTENDED key determines the display resolution of the two ECG
traces, whether or not both traces are currently displayed on the monitor.
Table 1: Display Resolution
The factory default setting for display resolution is monitor mode.
WARNING:
•
ECG detection circuitry may continue to count the
pacemaker rate during occurrences of cardiac arrest or
some arrhythmias. Do not rely entirely upon ECG rate
alarms. Keep pacemaker patients under close surveillance.
•
The system may insert pacemaker flags into the ECG
signal in response to signals that are not pacemaker
pulses. Therefore, if you use a Spacelabs Medical monitor
to observe pacemaker performance, you must take into
account all possible sources of pacemaker flags.
•
Use the pacemaker manufacturer's performance analyzer
as the primary means of evaluating pacemaker operation.
!
•
RESTORE SETTINGS changes the user-configurable options
for
all
parameters in the module.
Key
Display Resolution
Monitor
(0.5 – 30 Hz)
Extended
(0.05 – 30 Hz)
!
•
Changing the display resolution does not change the waveform
bandwidth used to analyze the ECG signals for the arrhythmia
and ST segment level.
To restore default settings:
1
Touch ECG
2
Touch SETUP
3
Touch RESTORE SETTINGS
4
Select YES
To change the display resolution:
1
Touch ECG
2
Touch SETUP
3
Select MONITOR or
EXTENDED