Entering Patient Information
The ADMIT/DISCHARGE menu enables you to enter a patient identification (ID)
number, name, height, weight, and body surface area (BSA).
Discharging a Patient
A patient is discharged by first removing the battery from the 91343/91347
Ultraview Digital Telemetry Transmitter. The monitor displays the squelch
waveform followed by the message INTERMITTANT SIGNAL LOSS after a short
delay. An alarm condition is displayed on the monitor because of the signal loss.
The message IS SIGNAL LOSS PERMANENT? appears with keys labeled YES
and NO in the waveform zone. Touch YES to indicate that the signal loss is
permanent. Touch NO to cancel the discharge operation.
The next message displayed is DISCHARGE THE PATIENT?. Touch YES to
continue the discharge process. Touch NO to cancel the discharge operation.
The monitor displays PURGES DATA-ARE YOU SURE? Touch YES to discharge
the patient and erase all patient data. The intermittent signal loss alarm is then
canceled. Touch NO to cancel the discharge operation and cause the message IS
SIGNAL LOSS PERMANENT? to appear in the waveform zone.
!
•
Admitting a new patient purges data from the previous patient
on that telemetry channel.
WARNING:
•
During INTERMITTANT SIGNAL LOSS message activation,
the display of SpO
2
and NIBP data is disabled.
To admit a patient:
1
Touch MONITOR SETUP
2
Touch ADMIT/DISCH
3
Select subnet (UCW and 1700
only)
4
Select bed/room number for
channel
5
Touch ADMIT
6
Select YES
7
Use keyboard to enter patient
info (UCW and 1700 only)
8
Select ID, NAME, HEIGHT,
WEIGHT, or BSA,
UV1050/1500 only)
9
Enter data using pop-up keypad
or keyboard ( UV1050/1500
only)
10
Touch ENTER
11
Repeat steps 7 - 10 until all data
has been entered
12
Touch ACCEPT (UCW and
1700 only)
To discharge a patient:
1
Remove battery
2
Disconnect the transmitter from
the patient
3
Select YES to confirm signal
loss permanent
4
Select YES to discharge
5
Select YES to purge data
Содержание Ultraview 91341
Страница 1: ...Exhibit H User Manual 2 FCC ID CM676A91341 600...
Страница 2: ...Ultraview Digital Telemetry 91341 91343 91347 Operations Manual 071 0774 00 Rev A...
Страница 20: ......
Страница 42: ......
Страница 60: ......