To restore the factory defaults for a configuration
Note:
The Factory Default button is not available if the current configuration is the
same as the factory default configuration.
1
Touch
Factory Default
, on the right side of the window.
The factory default settings are restored.
To rename a configuration
1
Touch the tab of the configuration that you want to change the name of. The
default names are
1
,
2
,
3
,
4
, and
5
.
2
Touch
Rename
on the right side of the window.
The Rename View window shows. The current name of the configuration
shows in the
Old Name
field.
3
Use the onscreen keyboard to type a new name.
4
Touch
Enter
.
Printer
There are two types:
• Bedside printer: A two-channel printer that provides automatic and manual
recordings of parameters in alarm conditions, requested waveforms, and non-
waveform data.
• System printer: A two- or four-channel printer that provides automatic and
manual recordings of parameters in alarm conditions, requested waveforms,
and non-waveform data. The system printer module can be used with either a
bedside or a central monitor.
You can configure the following settings on the
Printer
tab:
• Whether a recording is printed at a local or a network printer. Alarm recordings
can be printed at both locations.
• If a printer module is attached to the monitor, whether the printer module is
configured as a one-, two-, or four-channel printer.
• Duration of non-alarm recordings (12 or 20 seconds).
• Format of the Print Vitals report. Refer to
on page 8-5 for
additional details.
Note:
Duration settings are not enabled on all monitors. To access these settings,
contact your system administrator.