4
SPECIFICATIONS
Temperature range
D:S
-50 to 650 °C (-58°F ~ 1202°F)
12:1
Display resolution 0.1 °C (0.1°F)
<1000
1°F
>1000
Accuracy for targets:
Assumes ambient operating temperature of 23 to 25°C(73 to 77°F)
-50 ~ 20°C (-58°F ~ 68°F)
±2.5°C (4.5°F)
20°C ~300°C (68°F ~572°F)
±1.0% ±1.0°C (1.8°F)
300°C ~650°C (572°F ~ 1202°F)
±1.5%
Repeatability
-50~20°C (-58~68°F)
:
±1.3°C (2.3°F)
20~650°C (68~1202°F)
:
±0.5% or ±0.5°C (0.9°F)
Response time:
150ms
Spectral response:
8~14um
Emissivity Digitally adjustable from: 0.10 to 1.0
Over range indication LCD will show: “----”
Polarity: Automatic (no indication for positive polarity)
Minus (-) sign for negative polarity
Diode laser output: <1mW, Wavelength 630~670nm, Class 2 laser product
Operating temp.
0 to 50°C (32 to 122°F)
Storage temp. –10 to 60°C (14 to 140°F)
Relative humidity:
10%~90%RH operating, <80%RH storage
Power supply: 9V battery, NEDA 1604A or IEC 6LR61 or equivalent
Safety: CE Certificated, EMC tested and compliant