Voltage
Safety Rating
AC Frequency Range
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
Drop Protection
Safety Rating
NCV Detection Range
Battery Type
6-600V AC/DC
CAT IV 600 V, class 2, Double Insulation
50 Hz/60 Hz
32 °F to 104 °F(0 °C to 40 °C )
14 °F to 122 °F(-10 °C to 50 °C )
Max 85 % RH (NC)
≤6560 feet(2000 m )
6.5 feet (2 mètres )
CAT IV 600 V
100 V AC to 1000 V AC
2 - AAA, 1,5 V each
5. Touch the test lead probes to the component or wire you wish to check.
6. If the resistance is approximately 30
Ω
or less, a steady audible tone will
sound and a continuity connection symbol will appear
on the LCD screen. If the resistance is greater than 30
Ω
, the buzzer
will not sound and the continuity connection symbol will not appear
on the LCD screen.
Low Battery Indicator
When the battery voltage drops below approximately 2.7V,
the Low Battery Symbol will be displayed on the LCD screen
indicating that the batteries need to be replaced.
When this occurs, you should replace the batteries immediately.
If the battery voltage falls below 2.4V, the voltage tester will beep 5 times
and then immediately turn off.
CONTINUITY TEST MODE
GENERAL INFORMATION
When the battery voltage drops below approximately 2.7V,
the Low Battery Symbol will be displayed on the LCD screen
When this occurs, you should replace the batteries immediately.
If the battery voltage falls below 2.4V, the voltage tester will beep 5 times
Specifications
Replacing the Battery
WARNING:
To avoid electric shock, remove test leads from the voltage tester before removing
the battery cover.
1. Remove the one Phillips
screw from the battery cover
located on the back of the
41171N voltage tester.
2. Remove the battery cover.
3. Install two (2) new AAA
batteries.
4. Re-install the battery cover
and tighten the screw. Be
careful not to over tighten the
screw.
GENERAL INFORMATION