background image

Part 3 Operation

3.3 

Advanced Alarm Functions

The OMD-501D Trace Oxygen Analyzer is equiped with 2 

advanced relay contact alarms.  These alarms can be configured 

as normally open or normally closed, ha

s

 a delay model built-in 

and can be enabled or disabled when necessary.  Using the color 

display, the alarms are also color coded.  When not engaged, 

the alarms are shown in GREEN, when engaged the alarms are 

visible in RED - As seen in the picture to the right.

The alarms are rated at 5A @ 230VAC.  If you are connecting to solenoid valves or a pump whose cur-

rent can fluctuate greatly, it is adviseable to use a slave relay to ensure no damage occurs to the analyzer.

Setting Alarm 1 and Alarm 2 is identical to begin

 

 G

o ahead and 

move the cursor to the alarm configuration you want to adjust using 

the UP / DOWN keys in the analyzer MENU and press ENTER.

To adjust the value setpoint of ALARM1 simply move the cursor up / 

down to select ADJUST ALARM 1 and press ENT.  You will be 

prom

p

ted

 

to select a PPM or Percent value based on your alarm set 

point.  Press

 

ENT once selected.  From here

,

 you can use the UP / 

DOWN arrows

 

to adjust each digit which is highlighted in red.  Simply 

hit the enter button to move the cursor over.

The alarm function allows the user to set the alarm to trigger the relay 

contacts as a HIGH alarm or as a LOW alarm.  Simply open the ALARM1

menu and SET HIGH / LOW to meet your process requirements.

The alarm has a built in DELAY mode.  This is a nice feature to use if you

are working on your process and do not want to have the alarm relay

contacts trip.  You will set this in a similar fashion to setting the alarm.

The DELAY mode is in minutes, so adjust the value accordingly.

The alarm can also be enabled / disabled should it be hooked up but 

not required.  Move the cursor using the UP / DOWN keys to the 

ENABLE / DISABLE and select the ENTER key.

The alarms can function as normally open or normally closed.  This 

selection will be choosen when hooking up the rear wiring.  It is not 

completed through the software.  See the wiring diagram in section 2.4

for more information.

If any questions about the ALARM functions should arise, please contact

the factory for assistance or technical questions.

Содержание OMD-501D

Страница 1: ...OMD 501D Trace Oxygen Analyzer 4045 E Guasti Rd 203 Ontario CA 91761 USA Web www sso2 com Phone 1 949 398 2879 E mail sales sso2 com OMD 501D Manual Rev 1 3 1_2015 Instruction Manual...

Страница 2: ...nnections 2 6 Installing the Oxygen Sensor using the H3 Flow Through Sensor Housing 2 7 Installing the Oxygen Sensor using the H1 KF40 Sensor Housing 3 1 Understanding the Controls and Their Operation...

Страница 3: ...Southland Sensing Ltd under a strict quality proceedure To understand how the oxygen analyzer functions it is important to understand a little bit of the sensor characteristics The active components...

Страница 4: ...ight Dimensions 1 4 DIN 96 x 96 x 65mm Enclosure Flush Mount ABS Classification General Purpose Temperature 0 50 deg C Alarms 2 Adjustable NO or NC Power 12 28 VDC or 100 240VAC Signal Output 4 20mA R...

Страница 5: ...he factory If damage to any portion of the new analyzer is present stop and report damage to the shipping company as well as the factory The analyzer is shipped with all materials needed to install an...

Страница 6: ...ts The OMD 501D is designed as an indoor unit For field installations please refer to the NEMA4X IP66 en closure option offered by Southland Sensing Ltd Refer to part 3 operation section of this instr...

Страница 7: ...sensor housing it is important to make sure the bottom portion of the sensor housing is mounted flat and the top portion of the sensor housing which is manufactured of nylon is at the top This will i...

Страница 8: ...e polarity of the power be connected in reverse it is recommended to check polarity before running power to the unit It is also recommended to add an external fuse to the power inlet the analyzer does...

Страница 9: ...rom its box With scissors open nitrogen purged packaging and remove the sensor Visually inspect sensor for damage if damaged notify the factory immediately Remove the shorting pins across the back of...

Страница 10: ...he sensor housing and set to the side Inspect O ring for cracking replace if necessary Always lube your Orings Remove the sensor from its box With scissors open nitrogen purged packaging and remove th...

Страница 11: ...tion or SPAN CAL To adjust set or use the delay mode of the two onboard user programmable Alarm Relay Contacts All features are programmable selectable through the MENU bottom The UP DOWN arrows will...

Страница 12: ...0 1000 ppm 3 VDC Range 4 0 1 4 VDC Range 5 0 25 5 VDC From the HOME screen press the MENU key and the display will indicate Use the UP DOWN keys to move the cursor to allow the user select AUTO RANGE...

Страница 13: ...ercent value based on your alarm set point Press ENT once selected From here you can use the UP DOWN arrows to adjust each digit which is highlighted in red Simply hit the enter button to move the cur...

Страница 14: ...can verify the signal matches with the following formula Signal Output mA Reading Full Scale Range x 16 4 For example if we are reading 500ppm on the 1000 ppm range Signal Output mA 500 1000 x 16 4 Si...

Страница 15: ...ts are firmly touching the gold pogo pins on the housing The H1 sensor housing alread has a hole at the bottom of the sensor that exposes the sensing surface to ambient air Let the reading stabilize f...

Страница 16: ...tood Consult the factory if any questions arise Note For a new sensor purging with a zero gas for 4 6 hours will help the low end stability and response Connect the gas sample line and set the pressur...

Страница 17: ...he sensor No tools are required to replace the sensor Simply unscrew Counter Clockwise the collar H3 flow through sensor housing or top portion H1 KF 40 sensor housing Once free open the top portion o...

Страница 18: ...ide or Acetic Acid electrolyte Lead anode Handling Rubber or latex gloves safety glasses Storage Indefinitely Physical Properties Boiling Point Range KOH 100 to 115 C or Acetic Acid 100 to 117 C Melti...

Страница 19: ...PEL 10 ppm TWA Lead OSHA PEL 05 mg cubic meter Ingestion Electrolyte could be harmful or fatal if swallowed KOH Oral LD50 RAT 2433 mg kg or Acetic Acid Oral LD50 RAT 6620 mg kg Eye Electrolyte is cor...

Страница 20: ...pm 0 100ppm 0 1000ppm 0 1 0 25 Power 12 28 V DC 100 240 V AC Alarms 2 Adjustable High Low with Delay Mode Analog Output 4 20mA Isolated Range ID 1 5VDC in 1VDC increments Display Color with Backlight...

Отзывы: