Part 2 Installation
2.1
Receiving your New Oxygen Analyzer
As soon as you receive your new Portable Percent Oxygen Analyzer, carefully unpack the unit and acces-
sories and inspect the electronics module, sensor housing for damage and also verify the oxygen sensor is
present.
CAUTION:
Do not open the oxygen sensor packaging at this time. It is packed
in a Nitrogen purged bag. It is recommended that you read through the instruction
manual installation and operation sections before attempting to open the bag the oxy-
gen sensor is packed in. For questions, please contact the factory.
If damage to any portion of the new analyzer is present, stop and report damage to the shipping company
as well as the factory
The analyzer is shipped with all materials needed to install and prepare the system for operation. In some
instances, added sample system components are necessary to condition the gas sample before entering
the sensor housing. Southland Sensing offers free application consultation, and we encourage you to take
advantage of our engineers and their expertise.
Remote sensor housing installation: The sensor housing can be installed up to 6 feet from the electronics
module. Further than 6 feet will cause signal degredation which can cause the unit to give erronous read-
ings. It is also important to be mindful of EMI / RFI noise interference. Protection from EMI / RFI noise is
important for accurate readings.