Part 2 Installation

2.5 

Installing the Oxygen Sensor using H3 Flow Through Sensor Housing

The OMD-480 can accept either a PO2-160 or PO2-24 (CO2 Applications) percent oxygen sensor.  For 

proper selection with your application, please view our online application guide or contact your local sales 

rep.

Connect the process gas line and set your flow between 0.5 - 

5

.0 SCFH.

To Install the Sensor:

-  Remove the cell holder cap by unscrewing the stainless steel collar. 

-  Lift up on the top of the sensor housing and set to the side.

-  Inspect O’ring for cracking, replace if necessary.  

Always lube your Orings!

-  Remove the sensor from its box.  With scissors, open nitrogen purged packaging 

    and remove the sensor.

-  Visually inspect sensor for damage, if damaged notify the factory immediately.

-  Remove the shorting pins across the back of the sensor board contacts.

-  Place the sensor inside the housing with the metal screen mesh facing down and 

    the circuit board contacts facing up.

-  Return upper portion of the sensor housing to the stainless steel bottom.  Tighten 

    collar.  Hand tight is acceptable to create an airtight seal.

-  Start purge of process gas.

-  If the analyzer has not been calibrating, refer to section 4.1 for more information.

  

   CAUTION:  Prior to installing the oxygen sensor.  Read section 4.1 

on performing a span calibration.

H3 Sensor Housing

Top portion of sensor housing

Stainless Steel Collar

Bottom portion of sensor housing

Содержание OMD-480

Страница 1: ...OMD 480 Portable Percent Oxygen Analyzer 4045 E Guasti Rd 203 Ontario CA 91761 USA Web www sso2 com Phone 1 949 398 2879 E mail sales sso2 com OMD 480 Manual Rev 1 2 1_2015 Instruction Manual...

Страница 2: ...erview 2 3 Power Charging Connections 2 4 Gas Connections 2 5 Installing the Oxygen Sensor 3 1 Understanding the Controls and Their Operation 3 2 Measurement Range Overview 3 3 Setting the Day and Tim...

Страница 3: ...sed in the OMD 480 is designed and manufactured by Southland Sensing Ltd under a strict quality proceedure To understand how the oxygen analyzer functions it is important to understand a little bit of...

Страница 4: ...90 inch Display Full Color with Backlight Enclosure Rugged Polypropylene Flow Sensitivity 0 5 5 SCFH Gas Connections 1 8 Swagelok Tube Fitting Output Analog 0 1V DC Power Rechargeable Battery 100 240...

Страница 5: ...n of the new analyzer is present stop and report damage to the shipping company as well as the factory The analyzer is shipped with all materials needed to install and prepare the system for operation...

Страница 6: ...an OUTPUT connector for the 0 1V DC analog output and a CHARGE connector for the 12VDC power needed to re charge the battery On the right side of the panel is a mounted USB connector The OMD 480 can b...

Страница 7: ...d yellow means it is at the middle of its charge capacity and one bar and a red color and the unit should be recharged While charging the icon will have a AC Plug through the unit showing that the bat...

Страница 8: ...ensor from its box With scissors open nitrogen purged packaging and remove the sensor Visually inspect sensor for damage if damaged notify the factory immediately Remove the shorting pins across the b...

Страница 9: ...SPAN CAL Adjust the Date Time stamp on the display and to set the minutes interval when data logging via USB flash drive All features are programmable selectable through the MENU bottom The UP DOWN a...

Страница 10: ...display will indicate Use the UP DOWN keys to move the cursor to allow the user to select AUTO RANGE which will all ow the unit to cycle through all five ranges or MANUAL RANGE which will allow the u...

Страница 11: ...ne it is suggested to adjust the time and date to your local standard time To Set the Time simply select the MENU key and select the SET TIME function From here it will al low you to adjust the second...

Страница 12: ...ess enter and your interval will be saved For data logging purposes it is important to correctly set the date and time so that the time stamp will be saved along with the recorded oxygen value To set...

Страница 13: ...ze for about 1 minute and than proceed to the following steps in the OMD 480 menu SPAN Calibration To calibrate the analyzer press the MENU key use the UP DOWN keys to bring the cursor besides the opt...

Страница 14: ...the sensor should immediately be replaced Note make sure to read section 2 6 Installing the Oxygen Sensor before replacing the sensor No tools are required to replace the sensor Simply unscrew Counte...

Страница 15: ...ide or Acetic Acid electrolyte Lead anode Handling Rubber or latex gloves safety glasses Storage Indefinitely Physical Properties Boiling Point Range KOH 100 to 115 C or Acetic Acid 100 to 117 C Melti...

Страница 16: ...PEL 10 ppm TWA Lead OSHA PEL 05 mg cubic meter Ingestion Electrolyte could be harmful or fatal if swallowed KOH Oral LD50 RAT 2433 mg kg or Acetic Acid Oral LD50 RAT 6620 mg kg Eye Electrolyte is cor...

Страница 17: ...wer 12V DC Rechargeable Battery Analog Output 0 1V DC Data Loggin USB Flash Drive Display Graphical with Backlight Enclosures IP66 NEMA 4X Outdoor Enclosure Accessories Wall Mount Power Adapter 100 24...