13. DEVICE DESCRIPTION
─
Capacitor charge time,
─
Pacing threshold tests.
─
Data storage:
─
Therapy History Report,
─
Statistics (pace/sense, therapy, shocks, and battery voltage),
─
Up to 14 complete Holter records with event logs, marker channel notation, and
electrogram records.
The connector head has one port: ventricular bipolar pace/sense and RV & SVC
defibrillation coils.
The defibrillation port is compatible with the DF-1 standard.
Distal lead terminal connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the device via
feedthroughs.
Programming System:
The Sorin programmer is used in conjunction with specific programmer software to
interrogate and program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring:
INTENSIA VR 124 is also equipped with the RF wireless technology which enables to
remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.
28