6. CLINICAL STUDIES
Event Europe (N=38)
Number of events*
Number of patients
Percent of patients
Intent to treat but did
not
0
0
0.0
Non-device related
death
1
1
2.6
Explant
1
1
2.6
Complication
3
2
5.3
Observation
18
14
36.8
Serious non-related
other than death
12
7
18.4
Event All (N=60)
Number of events*
Number of patients
Percent of patients
Intent to treat but did
not
0
0
0.0
Non-device related
death
1
1
1.7
Explant
1
1
1.7
Complication
3
2
3.3
Observation
18
14
23.3
Serious non-related
other than death
13
8
13.3
* A patient can have more than one complication, observation, or serious adverse event, not
device-related.
Device Failures and Replacements:
No device failures or replacements occurred with Defender IV DR 612 during the study.
SORIN – INTENSIA VR 124 – U153A