4. Troubleshooting
CRANEX
®
Novus e
4.6 Bad quality images
Bad quality images can be due to one or more of the
following:
-
incorrect patient positioning
-
a badly aligned unit
-
the CCD sensor is not calibrated
Incorrect patient positioning
If the unit is producing bad quality images, fi rst make
sure that the user is positioning the patients correctly.
For information on how to position the patient refer to
the User's Manual.
A badly aligned unit
If patient positioning is correct, check the alignment of
the unit.
For information on how to align the unit refer to the In-
stallation and set-up manual.
The CCD sensor is not calibrated
Calibrate the CCD.
Refer to the section
Calibrating the CCD sensor
in the
installation manual.
Содержание Cranex Novus e
Страница 1: ...Service Manual for the CRANEX Novus e Dental X ray Unit 208710 rev 1 2013 01...
Страница 2: ......
Страница 23: ...14 Service Manual 2 Unit Description CRANEX Novus e 2 3 Unit dimensions...
Страница 34: ...CRANEX Novus e 2 Unit Description Service manual 25 2 7 Block diagram...
Страница 35: ...26 Service manual 2 Unit Description CRANEX Novus e 2 8 Fuse diagram...
Страница 40: ...CRANEX Novus e 3 Circuit Boards L1200 CPU Board Service manual 31 L1200 Block Diagram...
Страница 43: ...34 Service manual 3 Circuit Boards L1500 CCD Power Supply CRANEX Novus e L1500 Block diagram...
Страница 47: ...38 Service manual 3 Circuit Board L1600 Z Motor Driver CRANEX Novus e L1600 Block diagram...
Страница 51: ...42 Service manual 3 Circuit Boards L1700 Connector Board CRANEX Novus e L1700 Block Diagram...
Страница 54: ...CRANEX Novus e 3 Circuit Boards L1800 Generator Board Service manual 45 L1800 Block diagram...
Страница 58: ...CRANEX Novus e 3 Circuit Boards P1910 PFC Board Service manual 49 P1910 Block diagram...