Operating Manual | 3 Description of the device
SONO-PR 200
17 / 47
Revision: 2.0 | 2022-10-14
No. Type
Description/function
1 | 2 | 4 | 8 | 16 | 32 μJ
Available for “
SPIKE
” pulse only.
Voltage
Setting the pulser output voltage:
•
in “P/E” mode: 10 … 100 V (in 10 V increments)
•
in “T/R” mode and 50 Ω impedance: 10 … 100 V (in 10 V
increments)
•
in “T/R” mode and 1 kΩ impedance: 20 … 200 V (
in 20 V
increments)
Available for “RECT” pulse only.
LPF
Setting the low-pass filter:
200 MHz | 100 MHz | 50 MHz | 20 MHz
Gain
Setting the receiver signal gain in the following steps:
26 | 40 | 54 dB
Save
Saving all current settings as a specific setup.
Opens the “Setups” screen.
Impulse
Setting the pulse type:
•
RECT (unipolar-)
•
RECT (u) (optional)
•
RECT (bipolar) (optional)
•
SPIKE (peak)
Damping
Setting the pulser output impedance in
Ω
.
Attenuation
Setting the receiver signal attenuation:
Range
: 0 … 65
.5 dB
Increment: 0.5 | 6 dB
Bursts
Setting the number of pulse repetitions:
Range: 1 … 10 impulses
Available for “RECT” pulse only.
Recall
Loading saved settings.
Opens the “Setups” screen.
PRF
Setting the pulse repetition frequency (frequency of generated pulses)
for internal signal trigger:
Range
: 10 Hz … 20 kHz
Increment: 10 | 100 | 1000 Hz
HPF
Setting the high-pass filter.
100 kHz | 1 MHz | 3 MHz | 10 MHz