Operating Manual
SONOFLOW IL.52 V2.0
13
Revision: 1.0; Date: 2018-10-26
5.3 Switching output
Parameter
Default setting
Further options
(via SONOFLOW Monitor)
Type PNP
Activated (Flow switch)
NPN | Push pull
Selection
L / H
(low values = L [0 V],
high values = H [24 V])
Inverted: H / L
(low values = H [24 V],
high values = L [0 V])
Measuring error
H
L
Device fault
H
L
Configuration
Flow Switch
Fix L | Fix H | Volume dosing
(Threshold Volume in ml) |
Pulse volume (Pulse volume
in ml)
Threshold Flow On
1500 ml/min
0 … 3 000 ml/min
Threshold Flow Off
1200 ml/min
0 … 3 000 ml/min
Totalized volume
Enable volume calculation
Enable negative volume
Table 5: Default settings for switching output
5.4 RS485 interface
Parameter
Default setting
Further options
(via SONOFLOW Monitor)
Protocol
Dialog (sensor response
on demand only)
Auto Type 11
(sensor transmits measuring
data cyclically; interval
adaptable)
Table 6: Default settings for RS485 interface
Further information you will find in th
e document “Concept of Serial Communication” for
SONOTEC Flow Sensors. Contact our Service.
Содержание 200 08 0001
Страница 2: ...This page has been deliberately left empty...
Страница 9: ...Ultrasonic Flow Sensor Type SONOFLOW IL 52 V2 0 Operating Manual...
Страница 12: ...SONOFLOW IL 52 V2 0 Operating Manual 4 Revision 1 0 Date 2018 10 26 This page has been deliberately left empty...
Страница 14: ...SONOFLOW IL 52 V2 0 Operating Manual 6 Revision 1 0 Date 2018 10 26 This page has been deliberately left empty...
Страница 27: ......
Страница 28: ......
Страница 29: ...This page has been deliberately left empty...