Chapter 7: Replacement Procedures 43
C
HAPTER
7
Replacement Procedures
7.1
Display Subassembly Replacement
before making any repairs.
7.1.1
Required Parts
Service Assembly, Display, C1.9 (P02081)
Service Assembly, Display, C1.9 Japanese (P02545)
The Display Assembly does not include a label. One of the following labels must be
ordered and installed on the display. It must match the label on the display that is
being removed.
Label “180PLUS” (P01991)
Label “SonoHeart PLUS” (P01992)
Label “180 Vet” (P01993)
Label “180 II” (P01994)
7.1.2
Required Tools and Materials
•
A #1 Phillips screwdriver, 7.0 in. (17.8 cm / 177.8 mm)
•
A torque driver, 2.0–10.0 in./lb. (0.23–1.1 newton meter)
•
An anti-static mat
•
A wrist grounding strap
CAUTION:
`
Always use correct ESD procedures. ESD damage is cumulative and may not be
noticeable at first. Initial ESD symptoms may be slightly degraded performance
or image quality.
Содержание C1.9 PLUS
Страница 1: ...C1 9 PLUS Ultrasound System Service Manual...
Страница 4: ...iv C1 9 PLUS Ultrasound System Service Manual...
Страница 12: ...4 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 18: ...10 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 38: ...30 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 50: ...42 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 72: ...64 C1 9 PLUS Ultrasound System Service Manual This page intentionally left blank...
Страница 88: ...80 C1 9 PLUS Ultrasound System Service Manual A 8 Accessory Cables 1 2 3 4 5 6 8 9 10 7 11...
Страница 92: ...84 C1 9 PLUS Ultrasound System Service Manual...
Страница 98: ...90 C1 9 PLUS Ultrasound System Service Manual...