4.  PRINCIPLES OF OPERATION 

The ultrasonic power supply transforms AC line power to a 20 kHz signal that drives a 
piezoelectric converter/transducer.  This electrical signal is converted by the transducer to a 
mechanical vibration due to the characteristics of the internal piezoelectric crystals.  
 
The vibration is amplified and transmitted down the length of the horn/probe where the tip 
longitudinally expands and contracts.  The distance the tip travels is dependent on the 
amplitude selected by the user through the keypad.  As you increase the amplitude setting the 
ultrasonic processing intensity will increase within your sample. 
 
In liquid, the rapid vibration of the tip causes cavitation - the formation and violent collapse of 
microscopic bubbles.  The collapse of thousands of cavitation bubbles releases tremendous 
energy in the cavitation field.  The erosion and shock effect of the collapse of the cavitation 
bubble is the primary mechanism of fluid processing. 
 
The probe tip diameter dictates the amount of sample that can be effectively processed.  Probes 
with smaller tip diameters deliver high intensity ultrasonic processing, but the energy is focused 
within a small, concentrated area.  Larger tip diameters can process larger volumes, but offer 
lower intensity.   
 
The choices of a power supply and horns/probes are matched to the volume, viscosity and other 
parameters of the particular application.  Horns/probes are available for both direct and indirect 
ultrasonic processing.  The frequently asked questions (FAQ) section has more information on 
this subject. 
 

Please consult with a product specialist for assistance with selecting a probe for your 
application. 

 

Relationship of Amplitude and Wattage 

 
Ultrasonic processing power is measured in watts.  Amplitude is a measurement of the 
excursion of the tip of the probe (probe is also known as a horn).   

 

Some ultrasonic processors have a wattage display. During operation, the wattage displayed is 
the energy required to drive the radiating face of a probe, at that specific amplitude setting 
against a specific load, at that particular moment.  For example, the unit experiences a higher 
load when processing viscous samples than when compared to aqueous samples.   
 
The speed/cruise control on an automobile, can, to a certain extent, be compared to an 
ultrasonic processor.  The speed/cruise control is designed to ensure that the vehicle maintains 
a constant rate of travel.  As the terrain elevations change, so do the power requirements.  The 
cruise control senses these requirements and automatically adjusts the amount of power 
delivered by the engine in order to compensate for these ever changing conditions.  The greater 
the terrain rate of incline and greater the resistance to the movement of the vehicle, the greater 
the amount of power that will be delivered by the engine to overcome that resistance and 
maintain a constant speed. 
 
The ultrasonic processor was designed to deliver constant amplitude to your liquid sample 
regardless of these changes in load (much like the vehicle’s cruise control described above).  
As a liquid is processed, the load on the probe will vary due to changes in the liquid sample (i.e. 

Содержание VC505

Страница 1: ...Sonics Materials Inc 53 Church Hill Road Newtown CT 06470 USA Phone 203 270 4600 info sonics com Rev 10 21 www sonics com ULTRASONIC PROCESSOR Part No VC505 OPERATION MANUAL...

Страница 2: ...NNECTORS 10 5 4 CONVERTER CLAMPING 12 6 PREPARATION FOR USE 13 7 OPERATING INSTRUCTIONS 14 7 1 CAUTION 14 7 2 SETUP 15 7 3 OPERATION 16 8 TECHNIQUES FOR OPTIMIZING RESULTS 19 9 MAINTENANCE 21 9 1 PROB...

Страница 3: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Страница 4: ...ic Processor NEVER immerse the converter in liquids of any kind or let condensed moisture or liquid drip into the converter NEVER grasp an activated probe or touch the tip of a vibrating probe It can...

Страница 5: ...s Slow Blow 6 3 Amps Slow Blow Weight 16 lbs 6 8 Kg Dimensions 8 W x 15 25 L x 8 5 H 203 mm x 387 mm x 216 mm Output Voltage 1000 VRMS max Output Frequency 20 kHz Converter Weight 2 lbs 900 g Dimensio...

Страница 6: ...t configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Manufacturer is not responsible...

Страница 7: ...processing The frequently asked questions FAQ section has more information on this subject Please consult with a product specialist for assistance with selecting a probe for your application Relations...

Страница 8: ...e sample can readily be determined by visual observation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjusts...

Страница 9: ...29 Lab Stand with Clamp Sold separately 5 DESCRIPTIONS OF COMPONENTS FUNCTIONS OF CONTROLS 5 1 VC505 FRONT PANEL 5 2 VC505 REAR PANEL Power Supply Converter Cable Probe with Replaceable tip 630 0220...

Страница 10: ...ion from 1 second to 9 hours 59 minutes 59 seconds PULSER key Used with the numeric keys to set the pulse mode The ON cycle and OFF cycle can be set independently from 1 second to 59 seconds Red indic...

Страница 11: ...NNECTOR Pin No Description 1 Not connected 2 Not connected 3 Not connected 4 Enables connection to a frequency counter 5 Enables connection to an external power monitor 5 mv 1 watt 6 Ground 7 Energize...

Страница 12: ...NG Improper clamping can damage the system and void the warranty Using a sound enclosure part 830 00427 or stand part 830 00459 will ensure a proper fit Proper Clamping Improper Clamping Sound Enclosu...

Страница 13: ...e shipment ELECTRICAL REQUIREMENTS The Ultrasonic Processor requires a fused single phase 3 terminal grounding type electrical outlet capable of supplying 50 60 Hz at 100 volts 115 volts 220 volts or...

Страница 14: ...y twisting the converter housing as this may damage the electrical connections within the housing Overheating will damage the converter If continuous operation for more than 15 minutes is required see...

Страница 15: ...d with the unit 6 If you will be using a microtip or extender remove the flat tip on the end of the replaceable tip probe then attach the microtip or extender in its place 7 Microtips must be used in...

Страница 16: ...d the numeric keys for a 40 reading on the screen and then press the ENTER REVIEW key Pressing the AMPL key and the or key for a reading of 40 and then pressing the ENTER REVIEW key will also achieve...

Страница 17: ...rent from the elapsed time because the processing time function monitors and controls only the ON portion of the duty cycle For example for 30 minute processing time the elapsed time will be 1 hour if...

Страница 18: ...aying various operating parameters without process interruption Pressing the ENTER REVIEW key repeatedly during processing will consecutively display the following information a Selected amplitude e g...

Страница 19: ...larger than 50ml Microtips are high intensity and made for short processing times Using a microtip for long time periods will generate a considerable amount of heat Microtips should be used in pulse...

Страница 20: ...ive mixing you cannot ensure the entire sample volume will pass below the tip and become processed The probe should be submerged approximately halfway into the liquid but there are exceptions Before p...

Страница 21: ...PROBE MAINTENANCE Ultrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in contact wit...

Страница 22: ...o Converter Allow the alcohol to dry completely 4 Reattach probe to converter Screw the probe back onto the converter and tighten with the wrench set provided 5 The tips on replaceable tip probes can...

Страница 23: ...TEM CLEANING INSTRUCTIONS The power supply and converter may be cleaned using an acid free cleaning solution i e glass cleaner Probes should be cleaned using isopropyl alcohol Probes are made from tit...

Страница 24: ...been dropped A microtip being operated is damaged or worn past its useful life A fuse s has failed 10 1 OVERLOAD CONDITION If the Ultrasonic Processor stops working and an OVERLOAD indication is displ...

Страница 25: ...e adequate packing to insure against possible damage in shipment The Ultrasonic Processor should be sent to the address below with all transportation charges prepaid and return of shipment indicated R...

Страница 26: ...to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment listed above...

Страница 27: ...Each converter has 2 threaded ports for air cooling 10 psi of dry clean 5 micron filtered air is required One port is attached to the air source and the other port remains open as a vent Converter Co...

Страница 28: ...hose will exhaust the air from the converter outside of the enclosure 6 The exhaust hose only needs to exit the top of the enclosure by at approximately 8 inches or more COOLING AIR REGULATION AND AD...

Страница 29: ...of 29 Example of the air cooling hoses set up properly with the enclosure If you do not have a continuous duty air compressor we can offer a portable and reliable alternative Please contact us with qu...

Отзывы: