- 42 -
Problem
Possible Reason
What to do?
Measurement does not start
Pre-programmed
measurement does not
start.
Date or Time of
SOMNOscreen™
plus
and Date or
Time of PC are
different.
Check Date and Time on the PC. Insert the Compact
Flash Card (pre-programmed with a Manual Start) and
check the Date and Time on the
SOMNOscreen™
plus. Correct if necessary.
It is essential that the Date and Time is checked when
programming. Start a manual recording for 15 minutes
then abort the recording by pressing all 3 keys at the
same time. Transfer the Data and check it in the
Analysis Program.
Now initialize a Compact Flash Card for an Automatic
Start. Set the Start time to approximately 5 minutes
ahead of the current time and the recording duration to
15 minutes. Abort the recording by pressing all 3 keys
at the same time. Transfer the Data and check it in the
Analysis Program.
Defective snore
microphone
Replace the microphone. If a spare is not available, as
an alternative, use a nasal cannula if the measurement
is absolutely necessary. Change the montage and
start again recording.
Maximum
measurement period is
10 h.
Online Mode has
been activated.
In the Online mode, recordings are limited to 10h.
Deactivate the online mode to execute an offline
measurement. In offline mode recording periods up to
24h are possible (starting with Firmware version B11
and Software version 1.5.1).