Alert codes
Alerts are similar to faults but they do not reset the processor when they occur (i.e., the
device continues to deliver therapy pressure).
Device LED
Fault
LED
Error
Comments
Off
Flashes
15 times
Cannot regulate Device repeatedly detected the monitored pressure
outside of the therapy target range. The device
attempts to continue to deliver therapy. The alert is an
indication that filters and the mask should be inspected
for cleanliness and proper fit.
Filter Media change reminder
Every six months, the green LED flashes as a reminder that the Filter Media should be
changed. The LED only flashes when the CPAP is in Standby Mode. The CPAP will
continue to function normally. The reminder may be acknowledged in a manner similar to
acknowledging a fault (hold down the power button until the green LED stops flashing).
Acknowledging the reminder resets the reminder timer, that is it will flash again in six
months. The reminder should be acknowledged in conjunction with changing the filter.