Thymatron
®
System IV Instructions for Use
32
NOTE: Pressing the FlexDial™ will select the above program being displayed. Once a function or option is
selected with the FlexDial™ it remains in effect until changed, even when the unit is powered off.
To Exit FLEXDIAL™ Mode
:
There are two ways to exit FlexDial™ mode: Pressing the START/STOP button of the printer or pressing the
IMPEDANCE TEST button. The FlexDial™ dot L.E.D. will stop flashing upon exit.
1. Pressing the START/STOP button locks in the selection, generates a printed report of the 13 FlexDial™
selections in effect and exits the FlexDial™ mode.
2. Pressing the IMPEDANCE TEST button locks in the selections and exits the FlexDial™ mode without
generating a printed report.
FLEXDIAL™ Mode Procedure
For the remainder of this manual, selection of FlexDial™ functions and options will be shown by the following
shorthand notation with the explanation listed below:
FLEXDIAL™ [function] [options]
•
Press the FlexDial™ to display the most recently adjusted FlexDial™ function.
•
Rotate the FlexDial™ in either direction to display the desired function.
•
Press the FlexDial™ to flash-display the option in effect for that function.
•
Rotate the FlexDial™ to flash-display the other options.
•
Press the FlexDial™ to select the desired option and advance to the next option (if there is one) or to
return to FlexDial™ function level.
•
Press the IMPEDANCE TEST or START/STOP button to lock in the option and to exit the FlexDial™
mode.
Example:
FLEXDIAL™ CH 3-4 EMG-ECG, EEG-EEG means
1. Press the FlexDial™ to display the most recently-adjusted FlexDial™ function.
2. Rotate the FlexDial™ in either direction to display the “CH 3-4” function.
3. Press the FlexDial™ to flash-display the option in effect.
Содержание Thymatron System IV
Страница 57: ...Thymatron System IV Instructions for Use 56 ...
Страница 63: ...Thymatron System IV Instructions for Use 62 absorption and reflection from structures objects and people ...
Страница 79: ...Thymatron System IV Instructions for Use 78 ADDENDUM IV FLEXDIAL FLOW CHART ...