Thymatron
®
System IV Instructions for Use
25
SPECIFICATIONS
STIMULUS OUTPUT
Current: 0.9 amps constant current, limited to 450 volts, isolated from line current
Frequency: 10 to 70 Hz, in 10 Hz increments (to 140 Hz in LOW 0.25/0.3 Programs)
Pulsewidth: 0.25 to 1.5 ms, in 0.25 ms increments (0.3 ms is available)
Duration: 0.08 to 8.00 s in increments of equal charge
Maximum output: Standard maximum output across 220 ohms impedance, 504 mC (99.8 joules). Actual
(delivered) treatment output shown on printed report in mC
RECORDING
4 recording channels: channels 1 & 2, EEG; channel 3, EMG; channel 4, ECG.
8 user selectable gain positions for EEG channels (10, 20, 50, 100, 200, 500, 1,000 and 2000
μ
V/cm) and EMG
or ECG channels (50, 100, 200, 500, 1000, 2000, 5000 and 10,000
μ
V/cm)
REQUIREMENTS 100-130 volts (120 volts) AC, 60 Hz, single phase. 150 VA. (220-240 volt, 50/60 Hz,
switchable)
STIMULUS GENERATION
Waveform: bipolar, brief pulse, square wave
IMPEDANCE
Static Impedance Test: 0 to 3000 ohms static (+/- 100 ohms) at 800 Hz (L.E.D. and printed report)
Dynamic Impedance Measure: 0 - 500 ohms (printed report)
SEIZURE MONITORING
Channel specifications:
Maximum gain: EEG, 10
μ
V/cm; EMG 50
μ
V/cm; ECG 50
μ
V/cm
Common mode rejection: 80 dB
Isolation: full, opto-electronic
Printer paper speed: user selectable: 10 - 50 mm/s
Содержание Thymatron System IV
Страница 57: ...Thymatron System IV Instructions for Use 56 ...
Страница 63: ...Thymatron System IV Instructions for Use 62 absorption and reflection from structures objects and people ...
Страница 79: ...Thymatron System IV Instructions for Use 78 ADDENDUM IV FLEXDIAL FLOW CHART ...
Страница 80: ...Thymatron System IV Instructions for Use 79 ADDENDUM V MOCA TEST ...