INSTALLATION
Unpack the instrument, open the black case, and place the instrument on a firm, flat surface such as a hospital
cart. Check the instrument for any damage, and make sure the rear panel voltage designation matches that of the
electrical outlet to be used. Determine that all manuals and items are present. If required, have the Safety Check
performed as described in the Service Manual, Section 2.
Connect the power cable to the rear of the unit. Power the unit on and observe that the POWER switch light
comes on, the SELFTEST runs successfully, and “BASELINE” is displayed. Load a pad of thermal paper as per
the instructions inside the paper door (see page 12). Turn the power switch off and on again to verify that the
recorder prints the date and software version. This completes the installation.
OPERATING INSTRUCTIONS
The Thymatron® System IV features two front panel controls for display and selection of all treatment choices:
the PERCENT ENERGY stimulus dose dial and the FLEXDIAL™ function and option selector.
In addition, you will see a POWER switch (power on/off), an IMPEDANCE TEST button, a START/STOP
button (to manually control the 4-channel printer), a TREAT button (to deliver the treatment stimulus), two
alphanumeric L.E.D.s (the left one with 8 characters, the right one with 4 characters), and 5 individual dot
L.E.D.s (to indicate activation of the FlexDial™ selection mode, activation of the Safety Monitor alarm,
activation of monitoring ECG in channel 4, and whether the preset program or a user set selection is in effect).
POWER ON/OFF
Be sure the power cable is plugged into a grounded, 3-prong hospital-grade socket. Press the top half of the
front-panel POWER switch (labeled “I”) to turn the unit on; press the bottom half of the POWER switch (labeled
“0”) to turn the unit off.
SELF TEST
The Thymatron® System IV automatically tests the integrity of all circuits. When the unit is powered on, a
flashing nonsense symbol appears for several seconds in the 8-character L.E.D., followed by the flashing
message “SELFTEST” for a few seconds, then a self-test confirmation report is printed and the words “NO
BASE” appear, indicating that baseline EEG collection still has yet to be accomplished. (See IMPEDANCE
TEST procedure section for baseline EEG collection.)
The printed SELF TEST confirmation report will appear on the paper strip as:
Thymatron® System IV S/N [serial number here]
Date - Time S-IV version [software version] / [the number 60 or 50 Hz]
The Thymatron® S/N line can be replaced with the hospital’s name (see page 30).
Содержание Thymatron System IV
Страница 57: ...Thymatron System IV Instructions for Use 56 ...
Страница 63: ...Thymatron System IV Instructions for Use 62 absorption and reflection from structures objects and people ...
Страница 79: ...Thymatron System IV Instructions for Use 78 ADDENDUM IV FLEXDIAL FLOW CHART ...
Страница 80: ...Thymatron System IV Instructions for Use 79 ADDENDUM V MOCA TEST ...