Fraxel
®
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser System
Part Number 10-05170 Rev. O
Page 54 of 92
3.4.6
Do not allow the tip to lose contact with the skin during treatment.
3.4.7
The audible tones emitted by the unit during treatment are an
indication of hand velocity. If you are moving your hand at different
velocities, but you hear an unchanging monotonic sound, stop
treatment immediately. This indicates an IOTS fault. Stop treatment
and go to Simulate mode to test the IOTS. If the fault persists,
restart the system. If this fault condition persists, call Solta
Customer Service or your local authorized distributor of Solta
Medical lasers.
3.4.8
Never place the system in Treat mode and depress the footswitch
without intending to initiate a treatment.
3.4.9
Refer to the list of “Indications for Use” and “Contraindications”
before treatment.
3.4.10 Do not use visibly damaged (cracked, crazed, cloudy) tips.
Warning:
Do not move the footswitch while the laser is in Treat mode other than to initiate a treatment.
Do not move the footswitch by placing a foot in the housing and lifting it up. Do not pull the
footswitch by its cord. Do not move the laser by the footswitch cord. Do not place the
footswitch or cord in an exposed area where it may be damaged by for example, being
tripped over, or where a sharp object such as a scalpel may fall on it.
Warning:
Laser equipment not in use should be protected against unauthorized operation by removing
the key from the key switch. Keep the laser key in a designated place and allow only trained
personnel access to the key.
Warning:
Do not allow the tip to lose physical contact with the skin during treatment. Serious injury or
undertreatment may result if the IOTS enters a cycle of rapid starts and stops due to loss of
signal.
Warning:
Do not attempt to coil the umbilical beyond the resistance point of the armoring. Do not use
if the internal cables and wires are visible through holes or gaps in the umbilical armoring as
this may cause inadvertent exposure of laser energy to persons in the room.