Warning:
Contact lenses and spectacles do not provide adequate ocular protection. Do not substitute
for the specified eye protection. Failure to use appropriate eyewear with the Fraxel Laser
Systems increases the risk of injury to the cornea, sclera and soft tissue of the eye.
3.4 Operation of the Laser System
Warning:
This manual is not intended to be a complete guide to the use of the Fraxel 1550, Fraxel
1927 and Fraxel DUAL 1550/1927 Laser Systems. Solta recommends that all qualified
personnel who operate the laser system first seek training that includes, but is not limited to,
the following aspects of laser operation:
Laser
Safety
Laser/Tissue Interaction
Hands-On training by a qualified Solta representative
CAUTION: Solta does not and cannot recommend standard operating clinical
practice. The following guidelines are suggested as a starting point from which
operators may become familiar with good clinical laser technique through study of
the medical literature. The guidelines in ANSI Standards Z136.1, Z136.3, IEC60825-1
and EN207 are also recommended as reference material.
3.4.1
Check the laser prior to every procedure before admitting the patient
into the room.
3.4.2
Prepare the patient as directed with appropriate pain management,
and recommended contact lubricant (Refer to Section 4.8).
3.4.3
The footswitch should be controlled only by the operator to prevent
accidental firing of the laser.
3.4.4
The laser is designed to fire only when:
a) Either the 1550 nm or the 1927 nm laser is selected, and
b) The system is in “Treat” mode, and
c) The footswitch is depressed, and
d) The Intelligent Optical Tracking System (IOTS) senses
handpiece motion when the tip is in contact with the skin. See
Section 2.9 of this manual for a more detailed explanation of the
IOTS.
3.4.5
Press the footswitch only after positioning the laser handpiece over
the treatment area.