1.3. Checking the equipment at the receipt 

a)  At first,  check  the supply  conformity  with  the  ordered  equipment,  mainly  the  presence  of the 

parts necessary for the equipment mounting. Those to be used at the process will be given to 
the concerned department, for the installation preparation. 

b)  Place the sensor on a soft foam plate, connect it to processor (see §3.10) and switch on. The 

vibration  appears  at the rod,  the  viscosity  indication  is  close to  zero (or  close  to  the  minimal 
required  current  value  of  the  analog  output).  When  touching  the  rod,  the  information  has  to 
increase. 

In case of subnormal operation a) or b), check as follows : 

-

 

Power supply, connections, cables, 

-

 

The good condition of the vibrating rod (no bending on knock damages). 

Consult your distributor. 

1.4. Checking the equipment when placed at the process 

Before  filling  the  network  check  that  the  viscosity  information  is  stable  (vibrating  rod  in  air).  If 

not,  check  the  strength  of  the  sensor  fitting,  then  rotate  the  sensor  of  90°  (4  possible  positions). 
Choose the position where the information is the most stable. 

Locate this position, in order to restore it when the sensor is removed – put in place. 

Adjust the mounting offset, at room temperature. The rod is vibrating in air. 

When possible, note the viscosity information when a cleaning or rinsing solution is flowing. 

If the original calibration is convenient, one of the 2 above mentioned values can be taken as 

reference for periodic control of the equipment operation each time that the same conditions will occur 
(rod in air, or in the cleaning solution). Such an operation can be assimilated to a self-checking. 

If  the  original  calibration  has  been  modified  the  reference  values  will  be  of  course  those 

obtained with the new calibration. 

1.5. Periodic checking 

Conformity  to  regulations  relative  to  Quality  Insurance  implicates  a  periodic  control  of  the 

measuring  equipment  used  in  the  manufacturing  operations,  taking  in  consideration  (or  correcting) 
their drift in time. 

It is proved that this equipment drift is negligible. However, it is good to check their aspect and 

their response once a year, at the same time as the other process equipment. 

A  fast  test  is  many  times  available,  when  the  sensor  active  part  is  in  air,  or  immersed  in  a 

cleaning  or  rinsing  solution.    As  long  as  these  values  stay  similar,  we  can  say  that  the  sensor 
operation is right among its whole range (if no intermediate re-programming occurred). 

1.5.1. Offset adjustment in air 

The clean and dry rod is vibrating in air when the offset adjustment is carried out. 
 

The amplitude, corrected in offset, Vbrut, must be shift to the reference of 900 mV.  
See User Manual §5.1. 

Содержание MIVI 9610

Страница 1: ...Quality System certified instruments 9612 Version Technical Manual MIVI 9610 R f 254 31 0...

Страница 2: ...THE ROD MUST VIBRATE IN AIR 3 INSTALL THE SENSOR ON THE PROCESS AND FIX IT WITH ITS 4 SCREWS 4 REACH THE OFFSET ADJUSTMENT FUNCTION STARTING FROM THE BLOCK SETTINGS AND WITH THE SECURITY CODE 2 5 FOL...

Страница 3: ...2 2 Plane side mounting 10 2 2 3 Pot mounting for small flow rates or pilot plant 10 2 2 4 Special mountings 10 2 2 5 Replacement cap 10 2 3 PRACTICAL ADVICES 11 2 4 CHECKING 11 2 5 SENSOR WIRING 11 2...

Страница 4: ...E 23 6 2 RS485 INTERFACE 23 7 POWER SUPPLY 25 7 1 CHARACTERISTICS 25 7 2 SAFETY CONSIDERATIONS 25 7 3 POWER SUPPLY WIRING 26 8 GENERAL CABLING SCHEMATICS 27 9 SPECIFIC NOTES AND MANUFACTURING PARAMETE...

Страница 5: ...es UTILISATION Mode The UTILISATION Mode makes it possible to visualize in different forms viscosity and temperature values Visualization with bargraphs acquisition graph Access to the sensor identifi...

Страница 6: ...re The rod is vibrating in air When possible note the viscosity information when a cleaning or rinsing solution is flowing If the original calibration is convenient one of the 2 above mentioned values...

Страница 7: ...gas or II 2D dust Ex d IIC T1 to T6 gas Ex tD A21 IP67 T75 C T90 C T125 C T190 C T290 C T300 C dust Be sure the sensor s certification is in accordance with the security level required on your proces...

Страница 8: ...lded on the process line It means it can not be individually certified but with the whole process line 1 6 4 EMC and low voltage directive The processor 9601 is in agreement with EMC specifications de...

Страница 9: ...of the network or reactor If the fluid temperature varies widely and fast choose the upside down or horizontal position in order to allow a convenient air convection among the sensor body It is screw...

Страница 10: ...order to avoid parasitic vibrations the plate where the flange is welded must be thicker than 5 mm Advice preferably choose a sensor position with the rod to the top in order to have a total immersio...

Страница 11: ...to bend the vibrating rod A ring with the same dimensions of the protector s base must be placed on the head of the sensor in order to maintain the O ring IMPORTANT As soon as the sensor is removed s...

Страница 12: ...2 Standard and explosion proof models are both equipped with cable glands Those ones can match the ATEX norm depending on the MIVI version Earth terminal Pic 4 Shielding Cable Receiving coil Driving c...

Страница 13: ...NERAL PURPOSE Ex proof 200 C Std 200 C 135 mm 120 mm 100 mm 17 mm 30 mm Guard tube short slits Pot gas 156 mm 156 mm HIGH PRESSURE HP 200 C HP ex proof 200 C Pot gas Guard tube long slits Low pressure...

Страница 14: ...Technical Manual MIVI 9612 R f 254 30 0 14 2 7 Tightness O ring 29 82 x 2 62 76 or 79 70 or 99 O ring 21 95 x 1 78 O ring 30 00 x 2 00 O ring 30 00 x 2 50 Pic 6...

Страница 15: ...l dimensions Height 80mm Width 100mm Length 160mm Tightness IP66 Material Aluminium Temperature from 0 C to 45 C 3 2 Transducer s electronic board wiring The transducer s board is wired through the co...

Страница 16: ...26 mm Thickness 5mm Weight about 515 g Tightness IP 65 with panel mounting and IP20 on rear panel Operational temperature 0 C to 45 C 4 2 Display Display type STN LCD Illumination backlight LED yellow...

Страница 17: ...ply wiring 4 7 Connection blocks On the processor back module various categories are identified example Digital Inputs Analog Outputs We will call these categories connection blocks The diagram below...

Страница 18: ...oltage 0 10 V current 0 20 mA or 4 20 mA These inputs can be used with thermocouple or Pt100 On 9612 electronics these inputs are connected as 4 20mA and are specific to the plug of a densimeter Four...

Страница 19: ...om 02 to 07 and from 08 to 013 3 A max 250VAC or 30VDC Common 8 A max common plug not written see the connecting plan Minimum load 1mA 5VDC Lifespan 100 000 operations with a maximum load Response tim...

Страница 20: ...in If the user does not have the means to take in charge the wastes of our electronic equipments he should returned these equipments to our factory with a signed letter confirming that it is an equipm...

Страница 21: ...hole for the provided NC6 32 screw Scrape the panel s paint away to ensure the conductive connection Drive the screw into the hole and place the following items first washer ring cable shoe second was...

Страница 22: ...254 30 0 22 Step 4 Push the 4 mounting brackets into their slots on the processor sides shown as follow Step 5 Tighten the bracket screws on the panel Firmly hold the mounting brackets on the process...

Страница 23: ...on 4 RxD signal reception 5 0V reference 6 DSR signal 6 2 RS485 interface RS485 port is used to make the connection between the transmitter and the device Used COM port COM 3 Serial port type RJ 11 Ca...

Страница 24: ...munication capacity up to 32 devices Voltage limits from 7 V to 12 V as max differential Pin number Function 1 A signal red terminal 2 3 4 5 6 B signal white terminal Pins from 2 to 5 are not used for...

Страница 25: ...upply can be used provided that a 0V signal is connected to the frame If the device is mounted on a metal panel standard safety considerations require that the power supply must be earthed to avoid el...

Страница 26: ...the wire do not exceed a maximum torque of 0 5 N m We recommend that you use crimp terminals for wiring Use 26 14 AWG wire for all wiring purposes To maximize the system s performance electromagnetic...

Страница 27: ...cted to the 9612 device Pieces of advice The transducers must all have the same power supply Put a 120 resistance at the end of the line 1 4 W power is enough The RS485 shielding has to be connected t...

Страница 28: ...Technical Manual MIVI 9612 R f 254 30 0 28...

Страница 29: ...erature LOW 0 0 HIGH 999 9 HYST 0 0 HYST 0 0 Alarms attribution CURRENT OUTPUTS Physical value Current output value Viscosity LOW 0 0 4 HIGH 999999 9 20 BURN 4 Temperature LOW 0 0 4 HIGH 999 9 20 BURN...

Страница 30: ...999999 9 999999 9 Line 7 999999 9 999999 9 Line 8 999999 9 999999 9 Line 9 999999 9 999999 9 COMPENSATION TABLE X Y Line 0 0 0 0 0 Line 1 999 9 0 0 Line 2 999 9 0 0 Line 3 999 9 0 0 Line 4 999 9 0 0 L...

Страница 31: ...1 999 0 Line 2 999 0 Line 3 999 0 Line 4 999 0 Line 5 999 0 Line 6 999 0 Line 7 999 0 Line 8 999 0 Line 9 999 0 VISCOSITY ADJUSTMENT A 1 000 B 0 000 CALIBRATION DATA SERIAL N Y0 0 0000 a 1 0000 b 1 00...

Отзывы: