Technical Manual MIVI 9612
Réf. 254-30/0
28
Страница 1: ...Quality System certified instruments 9612 Version Technical Manual MIVI 9610 R f 254 31 0...
Страница 2: ...THE ROD MUST VIBRATE IN AIR 3 INSTALL THE SENSOR ON THE PROCESS AND FIX IT WITH ITS 4 SCREWS 4 REACH THE OFFSET ADJUSTMENT FUNCTION STARTING FROM THE BLOCK SETTINGS AND WITH THE SECURITY CODE 2 5 FOL...
Страница 3: ...2 2 Plane side mounting 10 2 2 3 Pot mounting for small flow rates or pilot plant 10 2 2 4 Special mountings 10 2 2 5 Replacement cap 10 2 3 PRACTICAL ADVICES 11 2 4 CHECKING 11 2 5 SENSOR WIRING 11 2...
Страница 4: ...E 23 6 2 RS485 INTERFACE 23 7 POWER SUPPLY 25 7 1 CHARACTERISTICS 25 7 2 SAFETY CONSIDERATIONS 25 7 3 POWER SUPPLY WIRING 26 8 GENERAL CABLING SCHEMATICS 27 9 SPECIFIC NOTES AND MANUFACTURING PARAMETE...
Страница 5: ...es UTILISATION Mode The UTILISATION Mode makes it possible to visualize in different forms viscosity and temperature values Visualization with bargraphs acquisition graph Access to the sensor identifi...
Страница 6: ...re The rod is vibrating in air When possible note the viscosity information when a cleaning or rinsing solution is flowing If the original calibration is convenient one of the 2 above mentioned values...
Страница 7: ...gas or II 2D dust Ex d IIC T1 to T6 gas Ex tD A21 IP67 T75 C T90 C T125 C T190 C T290 C T300 C dust Be sure the sensor s certification is in accordance with the security level required on your proces...
Страница 8: ...lded on the process line It means it can not be individually certified but with the whole process line 1 6 4 EMC and low voltage directive The processor 9601 is in agreement with EMC specifications de...
Страница 9: ...of the network or reactor If the fluid temperature varies widely and fast choose the upside down or horizontal position in order to allow a convenient air convection among the sensor body It is screw...
Страница 10: ...order to avoid parasitic vibrations the plate where the flange is welded must be thicker than 5 mm Advice preferably choose a sensor position with the rod to the top in order to have a total immersio...
Страница 11: ...to bend the vibrating rod A ring with the same dimensions of the protector s base must be placed on the head of the sensor in order to maintain the O ring IMPORTANT As soon as the sensor is removed s...
Страница 12: ...2 Standard and explosion proof models are both equipped with cable glands Those ones can match the ATEX norm depending on the MIVI version Earth terminal Pic 4 Shielding Cable Receiving coil Driving c...
Страница 13: ...NERAL PURPOSE Ex proof 200 C Std 200 C 135 mm 120 mm 100 mm 17 mm 30 mm Guard tube short slits Pot gas 156 mm 156 mm HIGH PRESSURE HP 200 C HP ex proof 200 C Pot gas Guard tube long slits Low pressure...
Страница 14: ...Technical Manual MIVI 9612 R f 254 30 0 14 2 7 Tightness O ring 29 82 x 2 62 76 or 79 70 or 99 O ring 21 95 x 1 78 O ring 30 00 x 2 00 O ring 30 00 x 2 50 Pic 6...
Страница 15: ...l dimensions Height 80mm Width 100mm Length 160mm Tightness IP66 Material Aluminium Temperature from 0 C to 45 C 3 2 Transducer s electronic board wiring The transducer s board is wired through the co...
Страница 16: ...26 mm Thickness 5mm Weight about 515 g Tightness IP 65 with panel mounting and IP20 on rear panel Operational temperature 0 C to 45 C 4 2 Display Display type STN LCD Illumination backlight LED yellow...
Страница 17: ...ply wiring 4 7 Connection blocks On the processor back module various categories are identified example Digital Inputs Analog Outputs We will call these categories connection blocks The diagram below...
Страница 18: ...oltage 0 10 V current 0 20 mA or 4 20 mA These inputs can be used with thermocouple or Pt100 On 9612 electronics these inputs are connected as 4 20mA and are specific to the plug of a densimeter Four...
Страница 19: ...om 02 to 07 and from 08 to 013 3 A max 250VAC or 30VDC Common 8 A max common plug not written see the connecting plan Minimum load 1mA 5VDC Lifespan 100 000 operations with a maximum load Response tim...
Страница 20: ...in If the user does not have the means to take in charge the wastes of our electronic equipments he should returned these equipments to our factory with a signed letter confirming that it is an equipm...
Страница 21: ...hole for the provided NC6 32 screw Scrape the panel s paint away to ensure the conductive connection Drive the screw into the hole and place the following items first washer ring cable shoe second was...
Страница 22: ...254 30 0 22 Step 4 Push the 4 mounting brackets into their slots on the processor sides shown as follow Step 5 Tighten the bracket screws on the panel Firmly hold the mounting brackets on the process...
Страница 23: ...on 4 RxD signal reception 5 0V reference 6 DSR signal 6 2 RS485 interface RS485 port is used to make the connection between the transmitter and the device Used COM port COM 3 Serial port type RJ 11 Ca...
Страница 24: ...munication capacity up to 32 devices Voltage limits from 7 V to 12 V as max differential Pin number Function 1 A signal red terminal 2 3 4 5 6 B signal white terminal Pins from 2 to 5 are not used for...
Страница 25: ...upply can be used provided that a 0V signal is connected to the frame If the device is mounted on a metal panel standard safety considerations require that the power supply must be earthed to avoid el...
Страница 26: ...the wire do not exceed a maximum torque of 0 5 N m We recommend that you use crimp terminals for wiring Use 26 14 AWG wire for all wiring purposes To maximize the system s performance electromagnetic...
Страница 27: ...cted to the 9612 device Pieces of advice The transducers must all have the same power supply Put a 120 resistance at the end of the line 1 4 W power is enough The RS485 shielding has to be connected t...
Страница 28: ...Technical Manual MIVI 9612 R f 254 30 0 28...
Страница 29: ...erature LOW 0 0 HIGH 999 9 HYST 0 0 HYST 0 0 Alarms attribution CURRENT OUTPUTS Physical value Current output value Viscosity LOW 0 0 4 HIGH 999999 9 20 BURN 4 Temperature LOW 0 0 4 HIGH 999 9 20 BURN...
Страница 30: ...999999 9 999999 9 Line 7 999999 9 999999 9 Line 8 999999 9 999999 9 Line 9 999999 9 999999 9 COMPENSATION TABLE X Y Line 0 0 0 0 0 Line 1 999 9 0 0 Line 2 999 9 0 0 Line 3 999 9 0 0 Line 4 999 9 0 0 L...
Страница 31: ...1 999 0 Line 2 999 0 Line 3 999 0 Line 4 999 0 Line 5 999 0 Line 6 999 0 Line 7 999 0 Line 8 999 0 Line 9 999 0 VISCOSITY ADJUSTMENT A 1 000 B 0 000 CALIBRATION DATA SERIAL N Y0 0 0000 a 1 0000 b 1 00...