2.2.1. Elbow mounting
•
The flange is welded on a right angle tee as indicated in
pic.1.
•
The minimal pipe diameter is of 32 mm.
•
The flange and the pipe axes have to be the same.
•
The flow direction is as indicated on fig.1 (unless for
fibrous fluids where the flow is inverted and the rod
protector removed. (See §2.2.5).
•
A free area of at least 150 mm length is necessary.
Advice : choose a sensor position in order to assure a
permanent fluid renewal and to avoid the existence of
“dead zone”.
2.2.2. Plane side mounting
•
The flange is welded on a metal plate as indicated on
pic.2.
•
The free area around the vibrating rod has to be at least
∅
40, 150 mm length.
•
In order to avoid parasitic vibrations, the plate where the
flange is welded must be thicker than 5 mm.
Advice : preferably choose a sensor position with the rod
to the top in order to have a total immersion of the
vibrating rod and to avoid the formation of “dead zone“.
2.2.3. Pot mounting, for small flow rates, or pilot plant
See picture 3.
2.2.4. Special mountings
The small sensor size allows numerous personalization
of its fitting according to the user's requirements. Consult
your distributor.
2.2.5. Replacement cap
Each mounting flange is provided with an obturation kit :
-
Cap and O-ring
-
4 fixing screws CHC M6 (or 8 fixing screws M8 for high
pressure sensor)
It allows the installation working when the sensor has to be
removed.