INTRODUCTION
Soehnle. Otherwise, danger may
arise for the user and the device.
•
Please inform Soehnle customer
service of unexpected operating
status or events.
•
Please do not place the device in
a location in which it is difficult to
disconnect from the power supply.
•
Keep the device outside of the
reach of small children/pets to
prevent the inhalation or swallo-
wing of small parts. The cable/cord
poses a risk of strangulation.
•
When stored at the minimum sto-
rage temperature between uses,
electrical medical devices must be
warmed up for at least 30 minutes
before they are operational. When
stored at the maximum permissib-
le storage temperature between
uses, electrical medical devices
must cool off for at least 30 minu-
tes before they are operational.
•
Please carefully monitor a possible
interruption of blood circulation
and a possible injury of the patient.
•
Do not place the cuff over a
wound as this may cause further
injuries.
•
Do not place the cuff on the same
arm on which other electrical
medical devices are placed.
•
Please ensure that the operation
of the device does not interrupt
blood circulation for a longer
period.
•
Avoid squeezing or constriction of
the connecting tube.
•
The Blood Pressure Monitor has
been tested pursuant to the requi-
rements of EN 1060-4:2004.
Metrological maintenance and care
information
•
Do not perform any metrological
maintenance if the device is in use.
•
For reliable operation, we recom-
mend that the device is subject to
metrological control once every
two years pursuant to Sec. 14
of the Medical Product Opera-
ting Directive (MPBetriebV). We
recommend testing the device’s
performance once every two years
and after metrological control and
repair, at a minimum by testing
the tolerance requirements of the
pressure display and tightness
of the cuff (testing at least at 50
mmHg and 200 mmHg).
•
The metrological control must be
performed by a licensed tester or
testing institute.
•
We recommend cleaning the de-
vice and cuff only with a soft cloth.
Storage information
•
If the device is not being used, store
the device in a dry place and protect
against heat, lint and fibres, dust and
direct sunlight. Do not place heavy
items on the storage box.
Information about handling batteries
•
Do not mix new and used bat-
teries.
Содержание SYSTO MONITOR 180
Страница 2: ...Deutsch 2 English 16 Fran ais 30 Italiano 44 Nederlands 58 Espa ol 72 Portugu s 86 100...
Страница 3: ...1 2 3cm 2 3 4 AAA Alkaline...
Страница 17: ......
Страница 31: ......
Страница 45: ......
Страница 59: ......
Страница 73: ......
Страница 87: ......
Страница 101: ......
Страница 102: ...ARAB 100 Soehnle 40 300 300 40 START STOPP APG AP 5 2009 10993 10 2010 10993 Soehnle 30 30...
Страница 103: ...ARAB 101 101 LCD 104 104 107 106 109 108 111 EEG 42 93 MDD 0123 SN BF EC REP...
Страница 105: ...ARAB 103 4 2004 1060 EN 14 MPBetriebV 200 50 EC 12 2008 EU 19 2012 0123 EEC 42 93...
Страница 106: ...ARAB 104 1 2 3 3 2 4 kPA mmHg 5 done LCD 60 1 2 2 1 3 20 5 1 2...
Страница 107: ...ARAB 105 104 7 5 1 104 0 133 1 106 1 1 2 2 LCD...
Страница 108: ...ARAB 106 ISH WHO 1999 110 180 HT 109 100 179 160 HT 99 90 159 140 HT 89 85 139 130 84 80 129 120 80 120 HT 25...
Страница 109: ...ARAB 107 1 2 2 3 4 3 1 2 1 AVG 2 3 2 1 1 1 1 3 2 60 1 dEL ALL 2 dEL done...
Страница 110: ...ARAB 108 1 E 2 E 3 E 11 E 10 E 20 E 21 E XX 01 02 EExx Out...
Страница 112: ...110...
Страница 114: ...Distributed by LEIFHEIT AG Leifheitstra e 1 56377 Nassau Germany www soehnle com 17763 A 2 2019...