18
■
Setting to zero
If the display does not read zero when
the scale is not loaded, press
.
Possible zero range 2%…+2% of the
weighing range
.
■
Tara calculate / deduct
Tare weighing
• Place the container on the platform.
• Press .
The display is set to zero.
The arrow sign points to
.
Weighing with manuel input tare
• Enter a known tare weight in kg.
• Press as long as the value is
flashing.
• Place the container on the platform.
• Place the weighing product on the
scale.
After the display stabilizes, the net weight of
the weighing product is displayed. The arrow sign
points to
.
Cancelling a tare
Press
when unloaded.
■
Counting
If piece weight is unknown
• Place the counted reference quantity
on the platform.
• Enter quantity and press
as
long as the number is flashing.
Further quantities of this part can be counted by
placing them on the platform.
If piece weight is known (manual input)
• Enter known piece weight in gram.
• Press as long as the number is
flashing.
Further quantities of this part can be
counted by placing them on the platform.
By small reference quantity or if the reference
weight is too small it will be shown on the display
panel “PIECE WEIGHT”.
Nevertheless the counting can be continued.
■
Setting target weight
For weight value or counting value you
can set a upper and lower limit.
• Press .
• To switch between counting (Cnt)
and weighing (Net) press
.
• Press .
The display panel “WEIGHT” shows “Hi Cnt” or
“Hi Net” (upper target value).
• Enter quantity or target weight
in gram or delete with
.
• Press to take over the value.
The display panel “WEIGHT” shows “lo Cnt” or
“lo Net” (lower target value).
• Enter quantity or target weight in
gram or delete with
.
• Press to take over the value
and to finish the input.
4. WEIGHING AND COUNTING
ENGLISH
Counting scale 9221