background image

 

 

supported.  If you have a VOLARO 

“WRIST ASSIST”, just hang the 

Wrist Assist 

from the “J” hook and do the 

same.    With  the  VOLARO  Stand,  the  person  being  transferred 

does  need  some  body  strength

.    We 

encourage anyone being lifted with the VOLARO Stand Lift to use their legs and arms as the lift is doing its 

work.  If you find that the person you are lifting is either less ambulatory than you originally thought, or the 

person is just not cooperative, use the VOLARO Full Body Lift Series 4 or Series 5.  

 

 

Transferring from Chair to Commode 

 

           After 

Applying  sling  to  Person

”  and  “Attaching  sling  to  Lift”  segments,  you  are  now  ready  to  lift. 

When the PA800C begins lifting and the abdomen becomes increasingly flat, snug the  security strap.  This 

will make the transfer much more comfortable for the person being transferred.   

 

 

Now  that  the  person  being  transferred  is  clear  of  chair,  unlock  the  brakes  and  roll  them  to  the 

commode or bathroom. Once in position, lock the brakes and attend to the person’s needs. Then lower them

 

down onto the seat.  

NOTE:  Be  sure  to  loosen  the  buckle  as  they  are  seated.

  If you wish, you may 

leave the lift with the person for added  support.   

 

 

Transferring from Chair to Bed 

 

           After 

Applying sling to Person

 and 

“Attaching sling to Lift”

 segments, you are now ready to lift. The 

VOLARO STAND can be used to transport a person back to their bed.  Lift the person to a standing position.  

Now  transport  the  person  to  the  side  of the  bed.    Center  them  and  make  sure  the  back  of  their  legs  are 

touching the bed.  Lower the person down until they are seated.  Be sure to leave slack in the loops. This will 

make it much easier for you to remove the colored loops. Unbuckle their legs from the shin rest safety strap, 

if  applicable.  Unbuckle  the  security  strap  around  their  abdomen  and  unhook  them  from  the  “J”  hooks.  

Remove the sling

. You may push the lift away. You may have to assist the person’s feet off

 the foot board. 

Be sure to use proper body mechanics by assisting the person to a lying position.   

 

Transferring from Bed to Chair

 

 

           When  transferring  from  bed  to  wheelchair,  it  is  required  that  the  person  be  assisted  to  a  seated 

position on the edge of the bed.  It is very important to use proper body mechanics.  In some cases you may 

wish  to  do  this  transfer  with  the  VOLARO  LIFT  Series  4  OR  the  VOLARO  LIFT  Series  5  and  then  use  the 

VOLARO STAND to go back to bed.  A person who has demonstrated to you their ability to stay seated on the 

edge  of  the  bed  without  help  may  not  need  much  assistance  with  proper  placement  of  the  sling.    For 

someone  with  little  ability,  apply  the  sling  while  they  are  in  a  lying  position  by  raising  the  shoulders  and 

inserting the sling behind the back.  Attach the security strap, then bring them to a seated position on the 

edge of the bed.  This will shorten the time you will have to support the person in a seated position before 

you hook them up.  

 

Standing and Ambulating 

 

           The VOLARO STAND-Series 5 allows a person to be lifted to a standing position with their feet flat on 

the floor. This provides an opportunity for the therapist to ambulate or perform standing exercises.  Some 

therapists utilize the VOLARO Seat Strap as an assist to help support the person while they are ambulating, 

standing, or resting.  

Lifting  a  person  with  their  feet  on  the  floor:

  After 

Applying sling to Person

 and 

“Attaching  sling  to  Lift”  segments,  open  the  legs  and  remove the  foot  board.      Place  the  sling

  behind  the 

person and hook them up, lock the brakes and adjust the shin rest.  Make sure their feet are located on a 

non-slippery  surface.  Adjust  the  shin  rest.  Hold  onto  the  lift  and  begin  to  lift  until  the  person  is  standing.  

Make sure they are straight and not leaning back before you begin ambulation. The “H” handles or optional 

ambulating bars  are  located  directly  in  front  and to  the  sides  of the patient  as  they  face the  lift. The bars 

work well when the person you are ambulating needs extra support.  If the person becomes fatigued, just 

lower them into a chair. 

Содержание Volaro Series 4 Stand

Страница 1: ...PA600 Operator s Manual SMT Health Systems 1380 Legion Road Detroit Lakes MN 56501 800 725 7761 www patientlift com Volaro Series 4 Stand...

Страница 2: ...er supplied with the lift Do not exceed the weight limit listed on the lift Do not use a sling that shows wear is torn bleached out or has loose threads Inspect sling before each lift PARTS SERVICE SM...

Страница 3: ..._____ _____________________________Date_____________ Assessment Care Planned by_________________________Date_____________ Can individual safely perform or 50 of lift transfer with only standby assista...

Страница 4: ...Diameter Screw Vertical Sliding Adjustable Shin Rest adjust ment range of 9 25 inches Horizontal Sliding Adjustable Shin Rest adjustment range of 4 inches 90 Degree Self Pivoting Shin Rest Steel Hand...

Страница 5: ......

Страница 6: ...ashing instructions are as follows Brush with warm soapy water and non chlorine disinfectant Rinse Drip dry only The sling rigidity makes it more difficult to machine wash Care should be taken if atte...

Страница 7: ...7...

Страница 8: ...ar down as you can Position the wings of the sling under the arms and make sure they are centered Secure buckle as low as possible around the lower abdomen Fasten the buckle and pull snug Tighten by h...

Страница 9: ...eet off the foot board Be sure to use proper body mechanics by assisting the person to a lying position Transferring from Bed to Chair When transferring from bed to wheelchair it is required that the...

Страница 10: ...ce to operate the lift This allows the operator to be next to the person being transferred and to operate the lift at the same time It also aids as a back up source to operate the lift Figure C Emerge...

Страница 11: ...cause it to overcharge If the power pac goes to GREEN in a short period it may mean you have a weak cell that will not take a full charge Plug power pac into charger every night Do not remove until ch...

Страница 12: ...VOLARO STAND PA600 PA600S PA600 EXPLODED VIEW...

Страница 13: ...PA600 PARTS LIST...

Страница 14: ...age them Circuit Board Wiring Diagram 1 Positive up to Switch Small Green 2 Positive Power Large Red 3 Negative Power Large White 4 Positive to Switch 2 Small Red 5 Safety Switch 2 Small White 6 Up Th...

Страница 15: ...ac 1 Power pac runs down quickly 2 Power pac light stays on 1 Power pac is worn out or not being charged when red charge light comes on lift 2 Power pac is worn out or circuit board is bad 1 Make sure...

Страница 16: ...ds 4 Get supplies ready sling wheelchair blanket or lap robe 5 Get help from experienced team member for this procedure 6 Have assistant stand on opposite side of bed to assist with transfer 7 Provide...

Страница 17: ...member to adjust the safety strap as the resident comes up to keep it snug at all times 20 Lift resident until buttocks have cleared the bed Give assurance to the resident 21 Pull the lift clear form...

Страница 18: ...Check the movement of the lift remove hair from the casters if needed Check leg covers if cracked replace with new ones Check all vinyl covers if worn replace with new ones Check all external fittings...

Страница 19: ...ut 8 Actuator Noise Up Down Replace bearings after 5 years Grease should be greased at least semi annually 9 Power Box Plastic Guides in Place Terminals Remote Jack Holder or Lens 10 Thumb Switches Le...

Страница 20: ...this form for Each Lift Date Checked By Repairs Needed Completed SMT Health Systems www patientlift com sales smthealthsystems com 800 725 7761 Lift Inspection must be done Monthly sling included Mod...

Страница 21: ...Motor Actuator Any Other Moving Parts Limited Warranty on all parts only under conditions of normal and intended use Main Structure Main Integrity and Welds of Frame Base Legs 2YEAR Limited Slings Ma...