56
S E C T I O N 5
M a i n t e n a n c e S o l u t i o n s
■
Switch off and disconnect the Unit.
■
Empty the unit once more.
■
Rinse reservoir with distilled water again.
■
Refill the unit with 1.4 litres of sterile distilled water.
■
Replace the Fill Port plug.
NOTE:
Unless using one of the approved maintenance solutions, use only
sterile distilled or de-ionised water. Failure to do so may lead to a
build-up of mineral deposits in the recirculating solution path which
may impair heater performance.
Changing Recirculating Solution (Isopropyl Alcohol)
Since the Isopropyl Alcohol solution gradually loses potency,
it must be replaced at the 30-day service interval using the
following procedure:
■
Prepare two batches of 1.4 litres of recirculating solution
according to formula 2 above for 35% Isopropyl Alcohol
maintenance solution.
■
Drain the HOTLINE®, and refill with one of the prepared
batches.
■
Install an L-70 or L-70 NI Disposable Administration Set in
the HOTLINE®’s Disposable Set connector.
■
Turn on the HOTLINE® and allow the fresh fluid to circulate
for 30 minutes.