
Chapter 4: Routine Maintenance
4-8
Surgivet
®
Universal CDS 9000 Operation Manual
Carefully pull out the glass Flowtube.
4.
Use a pair of thin Needle Nosed Pliers to remove the stop. (See Figures 4-11 and 4-12)
5.
Remove the ball from the tube and place it on a clean cloth.
6.
Figure 4-13: Remove the Ball
Squirt Isopropyl Alcohol into the tube.
7.
Gently tip the tube back and forth, allowing the alcohol to flow throughout the tube several times.
8.
NOTE! It may be necessary to clean flowtube with a plastic bristle brush or long Q-Tip.
Pour alcohol out of tube and properly dispose of used alcohol.
9.
Allow Flowtube to dry for about 10 minutes.
10.
Clean ball with isopropyl alcohol.
11.
Figure 4-11: Use a Needle Nosed Pliers
Figure 4-12: Remove the Stop
Stop
Содержание CDS 9000
Страница 2: ......
Страница 6: ...Table of Contents iv Surgivet CDS 9000 Operation Manual This page is intentionally left blank ...
Страница 12: ...Chapter 1 Introduction 1 4 Surgivet Universal CDS 9000 Operation Manual This page is intentionally left blank ...
Страница 43: ......
Страница 44: ...H Manufactured By Smiths Medical PM Inc N7W22025 Johnson Drive Waukesha WI 53186 1856 USA www surgivet com ...