Chapter 3: Machine Operation
3-4
Surgivet
®
Universal CDS 9000 Operation Manual
Getting Started
Check the machine to make sure it is operating normally and is leak free by performing a pressure test. A
pressure test is recommended to be performed daily. See Daily Machine Checklist on page 4-2.
Attaching the Universal F-circuit
Figure 3-2: Attaching the 32359B3 Universal F-circuit
WARNING! The Universal F-circuit is a rebreathing circuit that should only be used with animals that
weigh from 7 kg (15 lbs) or greater.
NOTE! When using the F-circuit, condensation will form in the outer tube. This will create warm fresh gas
which is good for the patient.
Locate the inhalation/exhalation valves on the front of the block.
1.
Attach the corrugated teal tube within the clear tube to the inhalation valve (left).
2.
Attach the short clear corrugated tube to the exhalation valve (right).
3.
Attach rebreathing bag to the bag port located underneath the exhalation valve.
4.
Connect to Exhalation Valve
Connect to Inhalation Valve
Patient End - Connect to
Endotracheal Tube or Mask
Connection for end tidal CO2 monitor
Exhalation Valve
Inhalation Valve
Block