Reservoir volume is zero. Pump stopped.
Alarm Priority High. The reservoir volume has
reached 0.0 mL. The pump will stop and not run. Select
Acknowledge to clear the alarm. Attach a new reservoir
and reset or edit the value of the reservoir volume,
if appropriate.
Upstream occlusion. Clear occlusion between
pump and reservoir.
Alarm Priority High. Fluid is not flowing from the fluid
container to the pump, which may be due to a kink,
closed clamp, or air bubble in the tubing between
the fluid container and pump. Delivery is paused and
will resume if the occlusion is removed. Remove the
obstruction to resume operation. The alarm will clear
when the occlusion is removed. You must acknowledge
this alarm after it clears if it has occurred and cleared
more than three times with 15 minutes.