Reports can be viewed at any time, with the pump
running or stopped.
1.
In the Tasks menu, press
or
to highlight View
Reports and press Select
2.
Press
or
to highlight
the desired report and
press Select
3.
Press Back to return to the
Select Report menu and
then Back again to return
to the Tasks Menu
Reports
100 mL
Stopped
Continuous
10:28 AM
8 Hour-Hydration
D5-0.45% NaCI 1000mL
mL
Back
Lock Keypad
Tasks
Press 'select' to begin
Display and Sound Settings
Change Time and Date
View Reports
View Advanced Tasks
Stop the pump if it is running.
In the Advanced Tasks menu,
press
or
to highlight
Patient Permissions and
press Select.
Priming security on/off
Setting this value to off allows
patients to prime the tubing
without having to enter a security code.
1.
In the Patient Permissions menu, press
or
to
highlight Priming Security On/Off and press Select
2.
Unlock the keypad
3.
Press
or
to set the security to on
(security code required) or off (no security
code required) and select Save
Delayed start security on/off
Setting this value to off allows patients to set delayed
starts without having to enter a security code.
4.
In the Patient Permissions menu, press
or
to highlight Delayed Start Security On/Off and
press Select
5.
Unlock the keypad
6.
Press
or
to highlight on (security code
required) or off (no security code required)
and select Save
Patient Permissions
100 mL
Stopped
Continuous
10:28 AM
8 Hour-Hydration
D5-0.45% NaCI 1000mL
mL
Back
Lock Keypad
Advanced Tasks
Press 'select' to begin
Patient Permissions
Air and Occlusions Settings
Alarm Settings
Security Settings