Fig. 6
Apply the Instrument
Secure the distal portion of the
assembly by placing the spring arms
of the Ankle Clamp around the ankle
proximal to the malleoli (Fig. 6).
Loosen the knob at the top of the
Distal Telescoping Rod and extend
the proximal portion of the assembly
to the joint line.
While applying a valgus stress to the
leg, introduce the paddle of the Distal
Femoral Resector into the affected
compartment (Fig. 7).
Be sure that the
paddle abuts the most distal aspect
of the femoral condyle, and that the
Tibial Resector engages the anterior
tibia. Then retighten the knob at the
top of the Distal Telescoping Rod.
Technique tip: If the Distal Femoral
Resector will not fit into the joint while
attached to the assembly, remove the
holding peg and resector. Then place
the Tibial Resector against the tibia,
and slide the Distal Femoral Resector
onto the dovetail with the paddle
inserted between the distal femoral
condyle and proximal tibial plateau.
Fig. 7