1. TIMER RESET -

 Pressing SELECT in this 

menu will reset the seconds counter in 
the Status Display to 0. 

2. QUICK SETUP -

 Pressing SELECT will 

initiate the Quick Setup sequence. 

3. DUAL MODE -

 Pressing SELECT will 

toggle the current setting of Dual Mode; 
enabling or disabling it. 

4. POWER LIMIT 

- Press SELECT to enter 

the power limit menu where the max 
power setting can be changed. Pressing 
Select in this menu will return the display 
back to Status Display. 

5. LIMIT MODE -

 In Power Limiting 

or Thermal Fold-back the power is 
reduced by changing the voltage or 
the frequency, while retaining the pulse 
width. This menu allow the selection of 
which method us used. Note: Frequency 
limiting is not possible with a DC Pulse. 

6. THERMAL FOLDBACK -

 If the PES

®

 gets 

too hot this mode, if enabled, will reduce 
the output power like Power Limit mode. 
Press SELECT in this menu to toggle the 
current setting of this mode: Enabling or 
Disabling it. 

7. LOW OUTPUT CURRENT -

 As a safety 

feature; If the PES

®

 output is activated 

but the unit does not detect any 
significant output current the output 
will be shut off and an error message 
displayed.  In this Menu the test for Low 
Output Current can be Disabled. Press 
SELECT in this menu to toggle this 
system setting; Disabling or Enabling it. 

8. EXTENDED TESTS -

 Pressing SELECT 

in this menu item will initiate a basic 
test of the PES

®

 unit. This should only 

be performed either by or under the 
direction of a engineer or technician. 

9. FACTORY DEFAULTS -

 Pressing SELECT 

in this menu item will set the current 
waveform and all stored waveforms to 
the Default Factory Setting. 

10. HISTORY -

 Press SELECT in this menu 

item to access the event history of the 
PES

®

 unit. Press the up or down arrow 

buttons to go forward or backward in 
history. Press SELECT or Menu to return 
to the Status Display 

SYSTEM MENU ITEMS ARE (IN ORDER): 

Содержание PES

Страница 1: ...10166 03 PES Manual USER S GUIDE...

Страница 2: ...a one year warranty Credit Refund Policy Customers returning equipment in new condition will be given credit to the applicable VISA Master Card account within five days from the date of the return A r...

Страница 3: ...be connected to an electrical outlet to charge the battery prior to use The PES is connected to the anesthesia tank via the anode and cathode cables The anode and cathode plates sit on either end of...

Страница 4: ...Recharge Input plug top right Check the indicator light on the external charging system to ensure that charge is being delivered bottom right Disconnect the charging system from the PES and AC outlet...

Страница 5: ...nu Timer Volts Duty Cycle Freq Pulse Type PORTABLE ELECTROANESTHESIA SYSTEM PES PORTABLE ELECTROANESTHESIA INSTRUCTIONS INSTRUCTIONS FOR CHARGING THE BATTERY Connect the external charging system to an...

Страница 6: ...cathode connectors firmly Connect anode and cathode cables to each electrode Plug in the remote switch SETTING UP YOUR PES HOOK UP DIRECTIONS 1 Ensure that the PES is OFF by pressing the red button an...

Страница 7: ...er time the electrodes accumulate a build up of oxidation making settings unpredictable and shortening battery life between recharges To clean the electrodes remove the protective screen covering the...

Страница 8: ...immersed in water Never use the PES if it is wet Dry unit off with a dry cloth before operating If necessary wipe control panel surface with a moist towel Solvents or cleansers may damage the control...

Страница 9: ...the unit is turned on a series of buttons and the centrally located display on the front panel of the device are used to interact with the PES and to program the electrical doses to be output OPERATIO...

Страница 10: ...d The output immediately decreases to zero upon release of the remote switch ON OFF UNIT IS ACTIVE WHEN LIGHT IS FLASHING 0 Off I On RECHARGE INPUT REMOTE OUTLET 26 4V DC 3 Amps TURN RIGHT TO ACTIVATE...

Страница 11: ...he period of the waveform is displayed Options for duty cycle vary between 1 and 99 After pressing DUTY CYCLE use the UP and DOWN to scroll through the available options Press SELECT to load the desir...

Страница 12: ...ary waveform timer disables the dual mode DEFAULT SETTINGS The PES comes with 10 sets of parameters pre programmed as default doses The individual parameters of the doses can be changed The 10 storage...

Страница 13: ...Primary waveform setting for that parameter Pressing SELECT whether or not a change is made will switch the display to the Secondary waveform setting PES MENU GUIDE 1 STATUS DISPLAY On Startup and at...

Страница 14: ...y for example and are in no specific order PES MENUS 6 VOLTS Pressing the VOLT menu button displays the current voltage setting The up and down arrows change the voltage value in the display and the S...

Страница 15: ...buttons limited by the currently set frequency and the PES s If Dual mode is enabled the first display shows the frequency setting of the Primary waveform If that waveform is a Burst of Pulses type t...

Страница 16: ...Cycle Status Display Primary Pulses Burst Primary Burst Duty Secondary Duty Cycle Status Display Primary Duty Cycle Secondary Pulses Burst Secondary Burst Duty Status Display Primary Pulses Burst Prim...

Страница 17: ...like Power Limit mode Press SELECT in this menu to toggle the current setting of this mode Enabling or Disabling it 7 LOW OUTPUT CURRENT As a safety feature If the PES output is activated but the unit...

Страница 18: ...SETTING 5 Standard 100 Volt Pulse of 100Hz with a 25 duty 100 volts for 25 milliseconds and 75 milliseconds off time 100 time each second SETTING 6 Burst of 3 100 Volt pulses of 250Hz with a 62 duty a...

Страница 19: ...functioning properly All personal protective equipment is in good condition All electrical connections in the system are clean and tight Electrodes are clean and in good condition with no loose or dam...

Страница 20: ...nd pulse type for 10 different waveforms Output Power 400W max continuous 16 000W peak max Overload Protection Excessive peak current average current or temperature will shutdown the unit before damag...

Страница 21: ...waveform the duty is multiplied by the Cycle Period but for a Burst Waveform the duty is multiplied by the Burst Period For the Standard wave the Pulse Width would be 0 25 times 200 milliseconds whic...

Страница 22: ...35 40 45 50 55 60 65 70 75 80 85 90 95 100 250 500 1K 5 10 15 2 20 3 25 4 2 30 5 2 35 6 3 2 40 7 3 2 45 8 4 2 2 50 9 4 3 2 55 10 5 3 2 2 60 10 5 3 2 2 65 10 6 4 3 2 2 70 10 6 4 3 2 2 75 10 7 4 3 2 2...

Страница 23: ...USER S GUIDE 21 2015 NOTES Use this section to write down settings for future use or other notes...

Страница 24: ...info smith root com 360 573 0202 Vancouver WA USA www smith root com Since 1964...

Отзывы: