PES: 

PORTABLE ELECTROANESTHESIA SYSTEM

8

www.smith-root.com

PULSE TYPE

The “PULSE TYPE” button provides 
options for type of electric current output 
by the PES

®

. After pressing PULSE TYPE 

use the 

 and 

 buttons to scroll through 

the choices. When the desired output is 
displayed on the screen, press SELECT to 
load the option into memory. 

Options for output:

 Ramped DC, Standard 

pulsed DC, and Bursts of pulses.

RAMPED-DC

This is when the output ramps up 
from zero to maximum (voltage level 
programmed by the user) over a period of 
1 second. This term applies to continuous 
DC only and not to pulsed waveforms. 
When using the dose TIMER, upon 
completion of the exposure, the ramped 
DC output will also decrease back to zero 
over a period of 1 second. Total exposure 
time includes the periods for ramping 
the output up and down; minimum total 
exposure time is 2 seconds.
When output is controlled manually (via 
remote switch), upon activation of the 
remote switch the output current starts 
low and takes 1 second to reach the 
selected maximum; if the output voltage 
was set for 100 volts the output would take 
1 second to raise the voltage to 100 volts 
level. The output will remain at this level 
until interrupted by release of the switch. 
Release of the remote switch deactivates 
the output immediately. 

Example:

 If the 

pulse timer is disabled when using Ramped 
DC, the DC output increases from 0 to 
the prescribed voltage level over a period 
of 1 second, when the remote switch is 
activated. The output will remain at this 
level until the remote switch is deactivated 
(released). The output immediately 
decreases to zero upon release of the 
remote switch.

ON/OFF

UNIT IS ACTIVE WHEN   

 

 

 LIGHT IS FLASHING

0-Off

I-On

RECHARGE INPUT

REMOTE OUTLET

26.4V DC

3 Amps

ANNUNCIATOR

ANODE 

CATHODE

A

D

J

U

S

T

TURN
RIGHT TO
ACTIVATE

PUSH OFF

Select

Save

Setup

Load

Setup

Menu

Timer

Volts

Duty

Cycle

Freq.

Pulse

Type

  PORTABLE

 ELECTROANESTHESIA 

SYSTEM

PES

PORTABLE ELECTROANESTHESIA INSTRUCTIONS

INSTRUCTIONS FOR CHARGING THE BATTERY

• Connect the external  charging system to the PES via the Recharge Input plug. 
• Connect the external charging system to an AC electrical outlet. 
• Check the indicator lights on the external charging system to ensure that charge is being delivered. 
• Disconnect the charging system from the PES and AC outlet after the battery is fully charged.

OPERATING INSTRUCTIONS 

The PES is designed for battery operation only. The PES should NOT be 
connected to an electrical outlet during operation. 
Red button – Device ON/OFF. For on, twist the button to the right. Press the 
button to turn the unit off.
LOAD SETUP – Review and select dose. 
Press LOAD SETUP, use the 

5

 and 

6

 buttons to review the electrical 

parameters to be applied. Press SELECT to load the settings for use. The dose 

to be delivered is programmed by the user and is defined by PULSE TYPE, 

TIME, VOLTS, FREQ, and DUTY CYCLE.

Use the  

5

 and 

6

 buttons to review the parameters available for use. Use 

SELECT to load settings into memory. 

Note: Any changes to the dose requires the user to review of ALL dose 
parameters (PULSE TYPE, TIME, VOLTS, FREQ, DUTY CYCLE). Use the  

5

 and 

6

 buttons to review the parameters and SELECT to load the parameters into 

memory. Display messages will prompt user for the required input. After all 
parameters have been reviewed, press SAVE SETUP, use the 

5

 and 

6

 buttons 

to select the memory location (page) where the newly defined dose will be 
saved, press SELECT. Up to 10 doses can be saved into memory.

ON/OFF

0-Off

I-On

RECHARGE INPUT

REMOTE SWITCH

& RECHARGE INPUT

TURN
RIGHT TO
ACTIVATE

PUSH OFF

PES® Control panel with central display

OPERATIONS

If the pulse timer is enabled, the output 
increases from 0 to the prescribed voltage 
level over a period of 1 second, remains 
at that level for exposure time minus 1 
second, then decreases to 0 over a period 
of the final second. 
If, however, the remote switch is released 
before the exposure-time has elapsed the 
output is immediately turned off.

STANDARD PULSED DC

Output immediately goes to maximum 
voltage and stays constant for the entire 
pulse duration, then drops to zero. This 
type of pulse is known as “rectangular”. 

BURST OF PULSES

Output is a series of dual-frequency 
rectangular pulses. This type of output may 
be referred to as dual-frequency pulsed 
DC, gated burst pulsed DC, or Complex 
Pulsed System. The pulses occur at two 
frequencies, an inner (burst) frequency and 
an outer (cycle).

Содержание PES

Страница 1: ...10166 03 PES Manual USER S GUIDE...

Страница 2: ...a one year warranty Credit Refund Policy Customers returning equipment in new condition will be given credit to the applicable VISA Master Card account within five days from the date of the return A r...

Страница 3: ...be connected to an electrical outlet to charge the battery prior to use The PES is connected to the anesthesia tank via the anode and cathode cables The anode and cathode plates sit on either end of...

Страница 4: ...Recharge Input plug top right Check the indicator light on the external charging system to ensure that charge is being delivered bottom right Disconnect the charging system from the PES and AC outlet...

Страница 5: ...nu Timer Volts Duty Cycle Freq Pulse Type PORTABLE ELECTROANESTHESIA SYSTEM PES PORTABLE ELECTROANESTHESIA INSTRUCTIONS INSTRUCTIONS FOR CHARGING THE BATTERY Connect the external charging system to an...

Страница 6: ...cathode connectors firmly Connect anode and cathode cables to each electrode Plug in the remote switch SETTING UP YOUR PES HOOK UP DIRECTIONS 1 Ensure that the PES is OFF by pressing the red button an...

Страница 7: ...er time the electrodes accumulate a build up of oxidation making settings unpredictable and shortening battery life between recharges To clean the electrodes remove the protective screen covering the...

Страница 8: ...immersed in water Never use the PES if it is wet Dry unit off with a dry cloth before operating If necessary wipe control panel surface with a moist towel Solvents or cleansers may damage the control...

Страница 9: ...the unit is turned on a series of buttons and the centrally located display on the front panel of the device are used to interact with the PES and to program the electrical doses to be output OPERATIO...

Страница 10: ...d The output immediately decreases to zero upon release of the remote switch ON OFF UNIT IS ACTIVE WHEN LIGHT IS FLASHING 0 Off I On RECHARGE INPUT REMOTE OUTLET 26 4V DC 3 Amps TURN RIGHT TO ACTIVATE...

Страница 11: ...he period of the waveform is displayed Options for duty cycle vary between 1 and 99 After pressing DUTY CYCLE use the UP and DOWN to scroll through the available options Press SELECT to load the desir...

Страница 12: ...ary waveform timer disables the dual mode DEFAULT SETTINGS The PES comes with 10 sets of parameters pre programmed as default doses The individual parameters of the doses can be changed The 10 storage...

Страница 13: ...Primary waveform setting for that parameter Pressing SELECT whether or not a change is made will switch the display to the Secondary waveform setting PES MENU GUIDE 1 STATUS DISPLAY On Startup and at...

Страница 14: ...y for example and are in no specific order PES MENUS 6 VOLTS Pressing the VOLT menu button displays the current voltage setting The up and down arrows change the voltage value in the display and the S...

Страница 15: ...buttons limited by the currently set frequency and the PES s If Dual mode is enabled the first display shows the frequency setting of the Primary waveform If that waveform is a Burst of Pulses type t...

Страница 16: ...Cycle Status Display Primary Pulses Burst Primary Burst Duty Secondary Duty Cycle Status Display Primary Duty Cycle Secondary Pulses Burst Secondary Burst Duty Status Display Primary Pulses Burst Prim...

Страница 17: ...like Power Limit mode Press SELECT in this menu to toggle the current setting of this mode Enabling or Disabling it 7 LOW OUTPUT CURRENT As a safety feature If the PES output is activated but the unit...

Страница 18: ...SETTING 5 Standard 100 Volt Pulse of 100Hz with a 25 duty 100 volts for 25 milliseconds and 75 milliseconds off time 100 time each second SETTING 6 Burst of 3 100 Volt pulses of 250Hz with a 62 duty a...

Страница 19: ...functioning properly All personal protective equipment is in good condition All electrical connections in the system are clean and tight Electrodes are clean and in good condition with no loose or dam...

Страница 20: ...nd pulse type for 10 different waveforms Output Power 400W max continuous 16 000W peak max Overload Protection Excessive peak current average current or temperature will shutdown the unit before damag...

Страница 21: ...waveform the duty is multiplied by the Cycle Period but for a Burst Waveform the duty is multiplied by the Burst Period For the Standard wave the Pulse Width would be 0 25 times 200 milliseconds whic...

Страница 22: ...35 40 45 50 55 60 65 70 75 80 85 90 95 100 250 500 1K 5 10 15 2 20 3 25 4 2 30 5 2 35 6 3 2 40 7 3 2 45 8 4 2 2 50 9 4 3 2 55 10 5 3 2 2 60 10 5 3 2 2 65 10 6 4 3 2 2 70 10 6 4 3 2 2 75 10 7 4 3 2 2...

Страница 23: ...USER S GUIDE 21 2015 NOTES Use this section to write down settings for future use or other notes...

Страница 24: ...info smith root com 360 573 0202 Vancouver WA USA www smith root com Since 1964...

Отзывы: