45
An extramedullary reamer alignment guide, which is
attached to the reamer, can be used to check the
position of the stem in axial alignment and depth ( the
end of the reamer alignment guide corresponds to the
tip of the reamer ).
Preparing the femoral stem connection
For reduction, femoral trials PS revision and tibial insert
trials PS are available.
See page 42 for the next step.
The largest reamer bushing «20» is inserted into
the femoral valgus adapter ( «4°», «6°» or «8°» ). The
Ø 20 mm drill is then drilled to the stop to prepare the
recess for the stem connection.
Important
Because the femoral stem connection for the PS femoral
component is deeper, the specific Ø 20 mm drill with PS
stop must be used.
Proceed with tibia preparation from page 31.